background image

U

SING

 

THE

 SurePower II Battery Pack

4

www.zoll.com

9650-000840-01 Rev. B

Displaying Battery Runtime

The 

SurePower II Battery Pack

can display its present status.

Note:

The 

SurePower II Battery Pack’s

 LEDs light 

only

 when the 

Display

 button is pressed.

To display the status of the SurePower II Battery Pack, press and 
release the 

Display

 button (momentary press) on the top of the 

battery.

When you press and release the 

Display 

button, all LEDs blink to 

indicate that they work. The battery then lights specific LEDs for 
approximately 10 seconds to indicate the battery’s runtime, calibration 
status, and fault state.

We describe what these LEDs indicate in the following sections.

SurePower II Battery Pack

 

Icon

LED Indications

This is the 

runtime indicator 

that displays the estimated 

runtime of the SurePower II Battery Pack in your defibrillator. 

When you momentarily press and release the 

Display

 

button on the SurePower II Battery Pack, the LEDs above 
these icons light for approximately 10 seconds to indicate 
the battery’s estimated run time. 

If the battery is fully charged, the LED over each of the five 
runtime indicator icons illuminates. As the battery 
discharges, the number of illuminated LEDs decreases.

As the SurePower II Battery Pack discharges, pressing the 

Display

 button will light, from left to right, the LED’s that 

indicate the battery’s run time. For example, if four LED’s 
light, the battery can power a defibrillator/monitor for at least
3 hours (displayed as 

3:00+

); if three LED’s light, the battery 

can power the defibrillator/monitor for at least 2 hours 
(displayed as 

2:00+

), and so on.

When you press the 

Display

 button on a battery that is 

installed and powering a defibrillator/monitor, the runtime 
indicator displays the run time for the battery in that specific 
unit.

When a SurePower II Battery Pack is not installed in a 
ZOLL defibrillator/monitor unit, the runtime indicator displays 
the runtime for the battery at a nominal drain rate for that
unit.

Note:

If the left-most LED of the runtime indicator blinks, 
the battery has less than fifteen minutes of use left, 
and defibrillator shutdown will soon occur. Recharge 
the battery as soon as possible.

Содержание SurePower II

Страница 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Страница 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Страница 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Страница 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Страница 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Страница 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Страница 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Страница 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Страница 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Страница 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...

Отзывы: