Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
29
IBP Test
Tools Needed
IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent
Setup
Connect the test cables to IBP patient simulator.
Do This...
Observe This...
Pass/Fail
1.
Connect the IBP patient simulator to the
Propaq P1 channel.
2.
Set the IBP patient simulator to 0 mmHg.
3.
On the Propaq M, press > IBP.
4.
On the Propaq M, Press > P1->0<-.
Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
o
o
o
o
5.
Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
o
o
6.
Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.
Содержание Propaq M
Страница 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Страница 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Страница 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Страница 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Страница 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Страница 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Страница 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Страница 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Страница 74: ...70 9650 0804 01 Rev A...
Страница 78: ...74 9650 0804 01 Rev A...