Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
25
Heart Rate Alarm Test
Tools Needed
ECG simulator
Do this...
Observe this...
Pass/Fail
1.
Press the power button on the
Propaq M to turn the unit on.
Connect the ECG leads to the
ECG
simulator
. Set the simulator to 120
BPM and press the Leads quick
access key
(
)
on the Propaq M
unit and select lead
II
.
II
displays next to the Leads quick
access key.
o
o
2.
On the Propaq M:
•
Press the Next quick
access key
(
) key to
access the second set of
quick access keys.
•
Press the Alarm
quick
access key
(
).
•
Press the
Limits
softkey.
The Alarm Settings menu displays.
o
o
3.
On the Propaq M, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:
•
Set the Upper Limit Enable
to ON.
•
Set the Lower Limit
Enable to ON.
•
Set the Lower limit to 30.
•
Set the Upper limit to 150.
Use the navigation keys to exit the
Alarm Settings menu.
4.
On the Propaq M, press the Alarm
Suspend key (
).
The red alarm suspension timer
displays at the top of the next to the
message area.
5.
Remove a lead wire from the
ECG
simulator
.
T
he
ECG LEAD FAULT
message
displays.
o
o
6.
On the Propaq M, press the Alarm
Suspend key (
).
The Alarm Suspend key flashes
.
7.
Reattach ECG Lead wire to
ECG
simulator
.
The
ECG LEAD FAULT
message no
longer displays.
o
o
Set simulator to 20 BPM or lower.
Heart Rate Value is highlighted and
the alarm tone sounds. The
HR/PR
LOW ALARM
message displays.
8.
On the Propaq M, press the Alarm
Suspend key (
).
The Alarm Suspend key flashes.
The
red alarm suspension timer displays
at the top of the screen next to the
message area.
o
o
9.
Set the simulator to 120 BPM.
I, II,
III...
Содержание Propaq M
Страница 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Страница 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Страница 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Страница 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Страница 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Страница 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Страница 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Страница 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Страница 74: ...70 9650 0804 01 Rev A...
Страница 78: ...74 9650 0804 01 Rev A...