CMCT Monitoring System
Instructions for Use
Page 17 of 31
5.4.2
CHEST LOCATION
1.
Stand in front of a mirror.
2.
Remove both parts of the Patch liner.
3.
Position the Patch on the upper left chest, just below the collarbone, and angle the
device towards the nipple. The arrow on the Patch should point up (↑). See
Figure 14
.
4.
Ensure the Patch is completely adhered to the skin.
5.
Stay still and wait for a green light to appear on the Sensor, which signals it is ready to
monitor (refer to
Section 6
Troubleshooting if a blinking amber light appears instead).
Note that the green light will last for several seconds, after which the light indicator will
remain off.
6.
Figure 15
illustrates examples of incorrect device placements for the front location.
Figure 8: Chest location incorrect placement
Figure 74: Device ( Patch) applied to front