ZOLL Base PowerCharger
4x4
Operator’s Manual
14
2SHUDWLQJ&RQWUROVDQG,QGLFDWRUV
The ZOLL Base PowerCharger
4x4
with AutoTest
automatically tests battery capacity with each battery
recharge.
The
BATT. READY
indicator light illuminates when the
battery is fully charged and capable of powering an
M Series equipped with Pace/Defibrillation and Pulse
Oximetry (SpO
2
) for approximately 1.5 hours in Monitor
mode. Fully charged batteries whose capacity is
insufficient will cause a
FAULT
light to illuminate.
With AutoTest, the full charging cycle is complete in
eight (8) hours or less for the PD 4410/Smart/
Smart Ready Battery and fifteen (15) hours or less for all
XL Batteries.
32:(5
Illuminates when the ZOLL Base PowerCharger
4x4
is
connected to live AC mains and the unit is ready for
charging and/or testing.
&+$5*(521
Illuminates when a battery is properly installed in the
corresponding charging/testing compartment and the
charge cycle is in progress.
Illuminates with the
TESTING
light during AutoTest and
recharge of the battery.
There are four
CHARGER ON
indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above it.
%$775($'<$XWR7HVW
Illuminates at the end of a test/charge cycle. Indicates
that the battery has been charged to 100% of its present
capacity and is ready for use.
Illuminates at the end of a battery test or battery
compartment test and remains lit as long as the battery
remains in the compartment. Indicates that the battery
has passed the capacity test and has sufficient capacity
to power an M Series equipped with Pace/Defibrillation
and Pulse Oximetry (SpO
2
) for approximately 1.5 hours
in Monitor mode, or that the battery compartment is
operating properly.
There are four
BATT. READY
indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.
)$8/7
Illuminates with the
TESTING
light when a battery fails
the capacity test. Also, illuminates when a battery
compartment or any other failure occurs.
There are four
FAULT
indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.
7(67,1*
Illuminates when either a battery capacity test cycle is in
progress or a battery compartment test is completed.
Illuminates with the
BATT. READY
light indicating a fully
charged battery, capable of powering an M Series
equipped with Pace/Defibrillation and Pulse
Oximetry (SpO
2
) for approximately 1.5 hours in Monitor
mode.
Illuminates with the
CHARGER ON
light during AutoTest
and recharge of the battery.
Illuminates with the
FAULT
light to indicate a battery that
has failed the capacity test.
There are four
TESTING
indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.
7(67EXWWRQ
Initiates battery test cycle or battery compartment test.
The battery capacity test cycle automatically charges,
discharges and recharges the battery.
If a battery is installed in the battery compartment and
the corresponding
TEST
button is pressed, a battery
capacity test is performed. If the battery compartment is
empty and the corresponding
TEST
button is pressed, a
compartment test is performed and an audible tone will
sound, indicating that the test has passed.
There are four
TEST
buttons. Each one corresponds to
the battery charging/testing compartment directly above.
6HWXS
Place the ZOLL Base PowerCharger
4x4
on a stable
secure surface. There must be at least four (4) inches of
clear space around the unit to allow for proper heat
dissipation during battery charging or testing. Do not
place anything on top of or beneath the unit, such as
blankets or cloths. Doing so may block the vents on the
unit, preventing proper dissipation of heat during
operation.
Turn the unit on by plugging the line cord into the power
module at the rear of the unit and then into live AC
mains.
%DWWHU\&KDUJLQJ3URFHGXUH$XWR7HVW
The ZOLL PD 4410/Smart/Smart Ready Battery may be
charged in any of the four battery compartments.
XL Batteries should only be charged in a ZOLL Base
PowerCharger
4x4
labeled “XL Battery Ready”. Each
battery charge compartment operates independently of