Appendix C. Technical Specifications
AutoPulse NXT User Guide
71
Electromagnetic Immunity Declaration (EID)
The battery charger is intended for use in the electromagnetic environment specified below.
RF Emissions CISPR
11
Class B
The platform and battery charger are suitable for use in all
establishments other than domestic and those directly con
-
nected to a low voltage power supply network which supplies
buildings used for domestic purposes, provided the following
warning is heeded.
WARNING.
This equipment is intended for use by healthcare
professionals only. This equipment may cause radio interfer
-
ence or may disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as re-orienting
or relocating the battery charger or shielding the location.
Harmonic Emis
-
sions IEC 61000-3-
2
Not applicable
Voltage Fluctua
-
tions / Flicker Emis
-
sions
IEC 61000-3-3
Not applicable
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.
Immunity test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment–guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
±15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV AC Mains
±1 kV I/O lines
5/50
100 kHz
±2 kV AC Mains
±1 kV I/O lines
5/50
100 kHz
Mains power should be
that of a typical commer
-
cial or hospital environ
-
ment
Surge
IEC 61000-4-5
±1 kV Line to Line
± 2 kV Line to Earth
±1 kV Line to Line
± 2 kV Line to Earth
Mains power should be
that of a typical commer
-
cial or hospital environ
-
ment
Voltage dips, short inter
-
ruptions, and voltage
variations on power sup
-
ply input lines
IEC 61000-4-11
>0% U
t
,for 0.5 cycle
1
At 0ºC, 45ºC, 90ºC,
135ºC, 180ºC, 225ºC,
270ºC, and 315ºC
>0% U
t
, for 0.5 cycle
At 0ºC, 45ºC, 90ºC,
135ºC, 180ºC, 225ºC,
270ºC, and 315ºC
Mains power should be
that of a typical commer
-
cial or hospital environ
-
ment.
If user requires contin
-
ued operation during
power mains interrup
-
tion, it is recommended
the battery charger be
powered from an inter
-
ruptible power supply
0% U
T
, 1 cycle and
70% U
T
, 25/30 cycles
Single phase at 0ºC
0% U
T
, 1 cycle and
70% U
T
, 25/30 cycles
Single phase at 0ºC
Voltage interruptions
0% U
T
, 250/300 cycles
0% U
T
, 250/300 cycles
Table C.7: Guidance and Manufacturer’s declaration –Electromagnetic immunity for the battery charger
Table C.6: Guidance and Manufacturer’s Declaration–Electromagnetic Emissions
Содержание AutoPulse NXT
Страница 1: ...14585 001 Rev 1 User Guide...
Страница 9: ...Chapter 1 Introduction 4 www zoll com This page intentionally left blank...
Страница 19: ...Chapter 3 Platform 14 www zoll com This page intentionally left blank...
Страница 27: ...Chapter 4 Band 22 www zoll com This page intentionally left blank...
Страница 31: ...Chapter 5 Battery 26 www zoll com This page intentionally left blank...
Страница 39: ...Chapter 6 Battery Charger 34 www zoll com This page intentionally left blank...
Страница 63: ...Chapter 9 Maintenance and Transport 58 www zoll com This page intentionally left blank...
Страница 67: ...Appendix A System Checklist 62 www zoll com This page intentionally left blank...
Страница 85: ...Quick Reference Guide 80 www zoll com This page intentionally left blank...