Chapter 3. Platform
AutoPulse NXT User Guide
13
To configure the default compression mode
The default compression mode is active when
the platform is turned on
. Configure the default
compression mode when you are setting up the platform.
1.
Ensure the platform is on.
2.
Check the default compression mode:
•
When the default compression mode is 30:2, Compression Mode is not illuminated.
•
When the default compression mode is Continuous, Compression Mode
button
is illuminated.
3.
T
o change the default compression mode, press
the
Compression Mode
button
until the platform beeps
once, then
release.
4.
Press and hold Compression Mode
button
. The platform beeps once. Hold for at least 4
seconds until
the
platform beeps again, then release.
Mute settings
Press Mute to silence ventilation, pause, and low battery beeps for 30 seconds.
If mute is activated, the beeps resume when:
•
30
seconds have passed.
•
Mute button is pressed again
.
Содержание AutoPulse NXT
Страница 1: ...14585 001 Rev 1 User Guide...
Страница 9: ...Chapter 1 Introduction 4 www zoll com This page intentionally left blank...
Страница 19: ...Chapter 3 Platform 14 www zoll com This page intentionally left blank...
Страница 27: ...Chapter 4 Band 22 www zoll com This page intentionally left blank...
Страница 31: ...Chapter 5 Battery 26 www zoll com This page intentionally left blank...
Страница 39: ...Chapter 6 Battery Charger 34 www zoll com This page intentionally left blank...
Страница 63: ...Chapter 9 Maintenance and Transport 58 www zoll com This page intentionally left blank...
Страница 67: ...Appendix A System Checklist 62 www zoll com This page intentionally left blank...
Страница 85: ...Quick Reference Guide 80 www zoll com This page intentionally left blank...