http://www.zketech.com
EBC-A01 User Manual
Copyright (C) 2006-2017 ZKETECH
3. Connecting patterns
3.1 Testing ports
The tester has four banana jacks which connect with A+/V+/V-/A-. A+ and A- ports are
connected with the positive and negative ends of the test objects, working as the current
channel in charging and discharging. V+ and V- ports are connected with the positive and
negative ends of the test objects, working as the voltage channel in voltage tests. This way of
using four individual cables will help eliminate the effect of wires on the test result and
improve the test precision.
3.2 PC connection
The tester can be connected with a computer through a designated USB-TTL cable (with Mini
port).
3.3 Input power:
The tester works on a power supply of DC 12V, 3.5A or above and interface specification of
5.5/2.1 (inside positive and outside negative).
4. Display and setting
4.1 Buttons
ON
—
start and stop
SET
—
set up and switch
INC
—
page up and increase
DEC
—
page down and decrease
4.2 Testing interface
The first line shows testing mode, voltage and current. Testing modes include:
CC
—
discharging at a constant current
CP
—
discharging at a constant power
CV
—
charging at a constant current and voltage
The second line shows working state (it shows OFF when the tester stops working and it turns
DSC
or
CHG
at
working.
It
reads
ATI
when
you
choose
automatic
charging-discharging-charging mode, and 1 refers to current working step), working time