http://www.zketech.com
EBC-A01 User Manual
Copyright (C) 2006-2017 ZKETECH
(minutes) and capacity (automatically switches between 0000mAh, 00.00Ah and 000.0Ah).
A short press on the
“
SET
”
button can switch between power (W) and capacity.
In the automatic
“
charging-discharging-charging
”
mode, you can check testing result of each
step with a short press on the
“▲”
or
“▼”
button:
Open the testing interface and press
“
ON
”
button to start testing and press again to stop it. In
the halted state press
“
SET
”
button for two seconds or longer to open the setting interface (this
function disabled when the tester connects to a computer).
4.3 Setting interface
The cursor sets automatically on the testing mode and you can move it to the option you want
to change. One press on the
“
SET
”
button to move one step to right. Press
“
INC
”
or
“
DEC
”
to
change the parameter. The parameter will change automatically if parameters set exceed
limitation. Press the
“
ON
”
button to switch to next setting option.
1) Discharging at a constant current
–
DSC-CC
The first line shows testing mode and discharging current.
The second line shows cut-off voltage (testing stops automatically when voltage reaches below
the set parameter) and the testing time (0-999 minutes, no limitation when it
’
s set 0).
2) Discharging at a constant power
–
DSC-CP
The first line shows testing mode and discharging power.
The second line shows cut-off voltage (testing stops automatically when voltage reaches below
the set parameter) and the testing time (0-999 minutes, no limitation when it
’
s set 0).
3) Charging at a constant current and voltage
–
DSC-CV
The first line shows testing mode and charging current.