Electrolysis Device SGM5EL Installation
20
If the amount of water vapor in the measuring gas is likely to cause a
condensation of water in a cold gas pipe, a water trap must be installed in
front of the gas inlet. (The penetration of water into the hot measuring cell
can destroy it and therefore must be absolutely prevented.)
4. If you want to install an external flow meter, you have to install it
behind
the gas outlet (where potential leaks do not have an influence on the
measuring result).
5. Connect the SGM5-EL with the power supply.
The measuring gas can also flow through the cold combi cell.
For using Swagelok®-fittings please note the manufacturer’s advice in the
appendix of this manual!
NOTE