Electrolysis Device SGM5EL Technical data
13
Modulation of the control parameter
r00
CR
R00 x.xxxxE+xx
CR
Read command variable
(values are read from the internal
memory)
R00 x.xxxxE+xx
CR
R00 x.xxxxE+xx
CR
Set command variable
0 - OFF
1 – Cell voltage [mV]
2 – Electrolysis current [mA]
3 – Oxygen concentration [ppm]
r01
CR
R01 x.xxxxE+xx
CR
Read setpoint
R01 x.xxxxE+xx
CR
R01 x.xxxxE+xx
CR
Set setpoint
r02
CR
R02 x.xxxxE+xx
CR
Read proportional value
R02 x.xxxxE+xx
CR
R02 x.xxxxE+xx
CR
Set proportional value
r03
CR
R03 x.xxxxE+xx
CR
Read reset time
R03 x.xxxxE+xx
CR
R03 x.xxxxE+xx
CR
Set reset time
r04
CR
R04 x.xxxxE+xx
CR
Read rate time
R04 x.xxxxE+xx
CR
R04 x.xxxxE+xx
CR
Set rate time
R!R!
R!
Save values
Error code
Meaning
error0
Transcription error
error11
Measuring temperature too low
error16
Thermocouple defective
Error20
System error
Error99
No connection to the SGM5-EL
Pin assignment of the socket SUB-D 9-pol. F
Pin assignment RS 232
Pin No.
Marking
2
TxD
3
RxD
5
GNDA