Manufacturer's EMC declaration
18
Page 33
Table 3:
Guidelines and manufacturer’s declaration –
Electromagnetic immunity
The Z
Lipo
Med
device is intended to be used in an electromagnetic environment as indicated below. The customer
or user of the Z
Lipo
Med
device must ensure that it is used in such an environment.
Immunity test
IEC 60601
–
test
level
Compliance level
Electromagnetic environment
–
Guidelines
Conducted disturbances by
HF fields according to IEC
61000-4-6
3 V
0.15 MHz
–
80 MHz
6 V in ISM band
between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
3 V
0.15 MHz
–
80 MHz
6 V in ISM band between
0.15 MHz and 80 MHz
80% AM at 1 kHz
In the environment of devices
which bear the following
symbol, interferences are
possible:
Radiated electromagnetic
HF fields according to IEC
61000-4-3
3 V/m
80 MHz
–
2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz
–
2.7 GHz
80% AM at 1 kHz