Technical information
9
Page 16
Basic device
power supply
220 - 240 Vac, 50 Hz
Fuse
2xT10AH, 250 V and 16 A automatic fuse in power switch
Sound power level
61.5 dB during normal operation (with operation of the cooling unit)
Power consumption
1200 VA
Protection class
I
IP class
IPX0 (device, applicator)
Vacuum
max. 693 hPa (520 mmHg)
Intensity levels
Intensities 1-10, 4 different pulse modes
Interface
USB
Maximum cooling
temperature on applied
component
-10 °C ± 5 °C
Dimensions
W 665 mm / D 629 mm / H 1082 mm
Weight
approx. 84 kg (with handpieces)
Operation
10 to 26 °C, 30 % to 75 % relative humidity, without condensation
at 700 to 1060 hPa
Storage / transport
(without filled water)
-10 to 50 °C, 10 % to 90 % relative humidity, without condensation
at 500 to 1060 hPa
Applicators
Medium handpiece
108 mm x 188 mm x 138 mm
Large handpiece
118 mm x 220 mm x 142 mm
Small handpiece
108 mm x 172 mm x 120 mm
Mini 360°-handpiece
119 mm x 139 mm x 89 mm
Small 360°-handpiece
100 mm x 169 mm x 100 mm
Medium 360°-handpiece
124 mm x 212 mm x 120 mm