Error messages
Troubleshooting
Disposal
28
Cable check
With stimulation current therapy in continuous current mode:
the patient current circuit is broken.
This message generally indicates failed electrodes, dirty electrode terminals or
a faulty patient fuse. Press OK to clear the message.
Excess current
This refers to an increase above the maximum allowable current. A current
increase in continuous mode generally indicates a fault in the device, while a
current increase in continuous voltage mode may occur because of a change in
the patient resistance (e.g., moist skin).
Press OK to clear the message.
If the error message appears again, contact customer service.