65
Ultrasound therapy
14.2 Description of the parameter
button
14
Selecting the mode
Press the two arrow buttons
to select the desired mode.
Modes:
•
continuous
•
pulsed with duty cycles of 1:2, 1:3, 1:5 and 1:10
Note:
The duty cycle of 1:10 only provides comfort heating.
Selecting the
frequency
Press the two arrow buttons
to select the desired frequency.
Frequencies: 20 Hz, 50 Hz and 100 Hz.
Water bath
Press the
button to switch the ultrasound head temperature monitoring to therapy in a
water bath.