Error messages, troubleshooting,
disposal
Error messages / Troubleshooting
15
Page 33
Malfunction
Possible cause
Remedy
No blood pressure
measurements available
Wrong time set
Check time on the display of
PremoPort
Four
and the PC. Correct
time on PC. (When transferring the
measurement plan, the time on the PC
is transmitted to PremoPort
Four
).
Wrong cuff type
Make sure that the cuff type used is
identical to the one set in the
measurement plan.
Leaky or defective cuff
Check the cuff and, if necessary, use a
new cuff.
Air circulation problems
Start a manual measurement and
check its progression.
Use “Receive and display” in
Diagnostik
Suite
together with the
measurement values to call up any
error codes from PremoPort
Four
and check the event table.