Technical information
Technical data / Storage and transport
7
Page 20
7.
Technical information
7.1
Technical data
Energy supply
Power supply
2 x 1.5 V AA alkaline batteries or
2 x 1.2 V AA NiCd or NiMH rechargeable
batteries.
Measurement cycles with fully
charged high-capacity
rechargeable batteries
250 – 300 individual measurements in 24 to
48 hours
Current consumption device off
Current consumption device on
Current consumption pump
process
Current consumption
measurement (without pumping)
1.2 mA
2.8 mA
< 1 A
120 mA
Power consumption
3 W max.
Display
Display
LCD
Signal transmission
Measurement data memory
Internal read-only memory
Data transmission
Optical
PC interface
Fibre optic adapter with a USB connector
Blood pressure
measurement
Blood pressure measurement
method
Oscillometric
Storage capacity
> 600 individual measurements
Pressure system
0-300 mmHg
Blood pressure measurement
range
30-260 mmHg
Heart rate range
40-200 bpm
Pressure sensor
Piezo-resistive
Pressurisation
Automatically controlled pressurisation
Safety
Maximum pressure: 300 mmHg,
independent overpressure valve
Pressurisation and quick-release
valve
Automatic pressure release valve
Design
Corresponds to EN 1060-1:1995
Electrical safety
Applied part
Type CF according to EN 60601-1
Protection against the
penetration of liquids
None
Operating mode
Continuous
Содержание PremoPortFour
Страница 1: ...Instructions for Use PremoPortFour EN ...