SECTION
Establish Femoral Alignment
TECHNIQUE TIP
1.B
If the epicondyles are visible, the epicondylar axis may be used as a
guide in setting the orientation of the Femoral Valgus Alignment Guide.
This does not set the rotation of the femoral component, but keeps the
distal cut oriented to the final component rotation. If desired, the guide
can be pinned to provide rotational stability
(Fig. 6)
.
2. Rotate
1. Retract
3. Release Retraction
Optional Pinning Holes
5° Right Valgus Setting
•
Retract the spring-loaded button on the top of the
Femoral Valgus Alignment Guide and rotate the valgus
adjustment dial to the desired valgus angle and release
the button to lock the valgus angle from 0° to 9°, ensuring
that the proper left or right setting is attained
(Fig. 4)
.
•
Place the Femoral Valgus Alignment Guide on the IM Rod
against the most prominent condyle and confirm the
desired valgus setting
(Fig. 5)
.
Fig. 4
Fig. 5
Fig. 6
75
mm x
3
.
2
mm
Trocar Tipped Drill
Pin (
2
.
5
mm hex)
00-5901-020-00
Femoral Valgus
Alignment Guide
00-5901-067-00
Pin/Screw
Inserter
00-5901-021-00
1