8
SECTION
4
1. Insert
2. Rotate
3. Extract
•
After final placement of the desired Cut Guide, insert
3.2mm trocar-tipped pins or 3.2mm Headed Screws (see
Screw Information section for examples) through the
oblique holes in the Cut Guide
(Fig. 23)
. Use a 1.27mm
(.050 in.) thick oscillating saw blade to complete the
anterior and posterior resections followed by the posterior
chamfer and anterior chamfer resections through the cut
slots.
(Fig. 24)
Upon completion of the cuts, use the Multi
Pin Puller or Pin/Screw Inserter to remove the oblique
pins/screws. Use the Persona Slaphammer to remove the
cut guide from the femur
(Fig. 22a & 22b)
.
Complete Femoral A/P and Chamfer Resections
TECHNIQUE TIP
4.C
If there is a risk of anterior notching, the 4-in-1 cut guide can be
removed, rotated 180° and be replaced on the distal femur. Holes can
then be drilled through the 2mm shift holes on the face of the 4-in-1
guide. The 4-in-1 guide then needs to be removed, rotated 180° and
be placed on the distal femur in the anteriorized holes. This will result
in a 2mm anterior shift of the 4-in-1 femoral resections. Verify that the
desired anterior and posterior resections will be attained.
Fig. 22a
Fig. 22b
Fig. 23
Fig. 24
TECHNIQUE TIP
4.E
- Areas where the bone transitions from high to low density, such as
from cancellous to cortical bone, can lead to the saw blade skiving
and inconsistent cuts. Use caution when resection in these regions.
Advancing an oscillating saw blade slowly through the bone while
using high frequency oscillations will help to ensure a flat, consistent
resection.
- Check the flatness of all bone resections to ensure sufficient contact
will result between the implant and the bone. If necessary, modify the
cut surfaces to be completely flat.
TECHNIQUE TIP
4.D
It is not recommended that the following Headed Screws are used
through the oblique holes of the Cut Guides, as the head of the screw
may interfere with the saw blade: 00-5791-041-00, 00-5791-043-00,
00-5791-044-00, 00-5061-063-00.
Multi Pin Puller
00-5901-022-00
75
mm x
3
.
2
mm
Trocar Tipped Drill
Pin (
2
.
5
mm hex)
00-5901-020-00
Pin/Screw
Inserter
00-5901-021-00