Instructions for use
3.2 Thermal threshold test
3
13
Implementation
The laser power is now set according to the user's best judgement and
delivered by pressing the footswitch.
During delivery of laser power, the elapsed time is displayed in the bar graph
(3) in the thermal threshold test window.
If the patient feels a sensation of heat, the laser delivery is stopped by pressing
the footswitch.
If the output lasts longer than 11 seconds, a message is displayed and the
power is increased by 0.5 W.
If the laser output is less than 7 seconds, a message is displayed and the
power is reduced by 0.5 W.
The thermal threshold test is now repeated with the modified power.
If the test is stopped between 7 and 11 seconds, the recommended power has
been determined. This is shown by a message.
The test mode is ended by selecting the "Heat threshold" window.
The specified laser power remains saved, and the original operating mode is
restored.