11  Contraindications 

The following list of contraindications, which is based on the current knowledge, has always 
to be observed when applying microwave treatment for therapeutic purposes! 
 
Always be sure to ask the patient about these contraindications, as not all contraindications 
are immediately recognizable by the therapist (e.g. pregnancy)! 
 
In  addition,  any  external  signs  that  might  point  to  the  existence  of  contraindications  (e.g. 
scars, etc.) should always be reason enough to ask the patient about contraindications! 
 
As a rule, any microwave therapy must be strictly based on an accurate diagnosis! 
 
 

11.1  Absolute contraindications 

-  Anesthetized patients 

-  Limitations in reactivity, cognitively impaired patients 

-  Febrile infections, acute infections, acute injuries 

-  Tuberculosis  

-  Haemorrhages or risk of haemorrhage  

-  Septic conditions and empyemas  

-  Malignant tumors and tumors that are not yet identified, active cancer treatment 

-  Patients with cancer in prior two years  

-  Severe heart diseases (heart valve diseases,  myocardial insufficiency,  myocardiac 

infarct, severe coronary sclerosis) 

-  Impaired or defective arterial circulation, ischemia, thrombosis 

-  Epiphyseal cartilages in children until completion of growth 

-  Lymphedema 

-  Severe osteoporosis 

-  Patients with pathologic conditions known to be sensitive to increased cell prolifera-

tion rates  

 

11.2  Local contraindications 

-  Metal and/or active implants including pumps, pacemakers and electrodes, metal in-

clusions, stents, stent grafts 

-  Swellings  

-  Thermal hyperesthesia  

-  Thermal hypoesthesia  

-  Acute inflammations, local infections  

-  Severe arterial obstructions (stage III and IV) 

Содержание MicroPro

Страница 1: ...Microwave Therapy Device MicroPro Instructions For Use...

Страница 2: ...any means electronic mechanical or otherwise without the prior written permis sion of Zimmer MedizinSysteme GmbH gbo Medizintechnik AG Part No 10105289 Version 1 4 Issue date 2022 03 09 Manufacturer g...

Страница 3: ...installation location 13 3 3 Transport of the unit 14 3 4 Unpacking the unit 14 3 5 Receiving inspection 14 3 6 Complaints 15 4 Description of the unit 16 4 1 Operator side 16 4 2 Patient side 17 4 3...

Страница 4: ...used therapy settings favorites 33 9 Maintenance 34 9 1 Legal requirements and regulations 35 9 2 Technical safety checks 35 9 3 Cleaning and disinfection 36 9 4 Repair 37 9 5 Disposal of the device a...

Страница 5: ...oPro microwave therapy unit from Zimmer MedizinSysteme GmbH 1 3 Intended use The MicroPro is used for microwave therapy diathermy By means of a high frequency electro magnetic field a tissue warming e...

Страница 6: ...r property damage are excluded if they are the result of one or more of the following causes Unauthorized use of the unit Improper operation and maintenance of the unit Operation of the unit with prot...

Страница 7: ...fies direct instructions for action This dash identifies enumerated items 1 7 Explanation of key symbols 0123 CE conformity sign Follow the instructions for use Danger for persons with pacemakers or s...

Страница 8: ...The catalogue number is placed adjacent to the symbol Indicates the manufacturer s serial number so that a specific medical device can be identified The serial number is placed ad jacent to the symbo...

Страница 9: ...out any work Please read and observe all safety instructions in this instruction for use 2 1 Responsibilities The microwave therapy unit may be used only by doctors physiotherapists or trained assist...

Страница 10: ...iator cable is not damaged the radiator cable is properly connected and screwed is not in contact with the patient or with conductive or energy absorbing objects only accessories cables radiator appro...

Страница 11: ...een the unit and power supply or data cables in walls ceilings and floors since the electromag netic radiation from the unit can pass these essentially without hindrance In selecting the location for...

Страница 12: ...outdoor operation Do not treat Patients on metal chairs or metal beds Keep credit cards camera mobile phone car key USB sticks memory cards and other interference sensitive storage media away from the...

Страница 13: ...t the mains plug is easily reachable to the operator Therefore you must read the safety instruction in Chapter 2 and the following infor mation By selecting a suitable location for setting up the unit...

Страница 14: ...tor and radiator cable it must be unpacked by at least two persons Proceed as follows Position the transport packaging so that the arrow mark is pointing upward Remove the safety bands from the transp...

Страница 15: ...diately Notification of the manufacturer and the remedy of the damage are generally carried out by the dealer Immediately create a damage report and send it to the transport company and to the dealer...

Страница 16: ...Medizintechnik AG Version 1 4 4 Description of the unit 4 1 Operator side Fig 4 1 Front view MicroPro Radiator Radiator cable Self locking plastic support arm extractable Display Touchscreen Housing...

Страница 17: ...MicroPro 17 gbo Medizintechnik AG Version 1 4 4 2 Patient side Fig 4 2 Back view MicroPro Mains fuse Appliance inlet Marking plate Connector for the radiator cable Mains switch...

Страница 18: ...Radiator cable Fig 4 4 Focus radiator Fig 4 4 Cradle radiator Fig 4 6 Long field radiator Fig 4 5 Circular field radiator Treatment side Treatment side Device Connection Radiator cable Device Connect...

Страница 19: ...d with different radiators in continuous or pulsed mode The application of high frequency energy in heat therapy has the advantage of greater depth penetration as opposed to simpler methods such as pa...

Страница 20: ...cal application areas are e g shoulder or knee joint jaw joint lower arm carpal region tenovaginitis etc The proper treatment distance between radiator and body part should be about 10 cm If the dista...

Страница 21: ...d at eyes Patients should wear radiation protection goggles 5 5 Microwave therapy with cradle radiator The cradle radiator enables a comprehensive exposure to radiation of body parts of medium and lar...

Страница 22: ...connector fully into the coupler and release the locking ring The locking ring must spring back to its original position Push and pull the radiator in the axial direction of the coupler to ensure that...

Страница 23: ...position with the holding arm pointing vertically upwards To overcome a threshold move the device slowly over it and if necessary lift the device slightly 6 3 Positioning the radiator Position the re...

Страница 24: ...output power According to Schliephake the patient s subjective sensitivity to heat can be divided into four stages according Dose I Heat not perceptible First set dose so that there is a slight sensat...

Страница 25: ...til the treat ment time has elapsed Remove the radiator from the patient and enquire the patient s condition 6 7 Placing out of operation In order to place the device out of operation no further measu...

Страница 26: ...yed out buttons In case of doubt the user can obtain further information via the direct help system The sliders or control wheels are operated as on a PC touch the slider and drag it to the desired po...

Страница 27: ...ts by moving the finger at the intensity regulator to the left Here an acceleration function and trailing is present allowing the intensity to be reduced in short time The USB port and the SD card rea...

Страница 28: ...can be seen at the top of the screen To the right is a flag indicating the currently set system language which can be changed by pressing it Next to it is the menu symbol Pressing the menu symbol ope...

Страница 29: ...call up the direct help press the button The button lights up green after selection Then press the button for which you desire information An information window opens which gives information about th...

Страница 30: ...system language Date Time Both the date and time as well as the date format and time format can be changed here Also it is possible to change the time zone Audio System tones such as the end of therap...

Страница 31: ...to set the impulse or continuous operation mode In pulse mode two further selection keys can be used to set the pulse frequency and pulse width In continuous oper ation the selection keys for pulse f...

Страница 32: ...1 Saving a therapy A therapy can be given a name and saved By opening the menu and then pressing the Save therapy as button a keyboard opens Enter the desired name for the therapy and save the therap...

Страница 33: ...o save the current settings of the therapy by pressing on a free favorite field on the right side of the screen Al already saved therapy under Own Therapies can be accepted and additionally added to t...

Страница 34: ...evice does not contain any parts that have to be maintained by the user Warning No parts of the device are allowed to be serviced while a patient is under treatment However the operator has the option...

Страница 35: ...e of the Medical Device Directive it is recommended to have the device checked by the manufacturer or by a maintenance service authorized by the manufacturer every 12 months Refer to separate periodic...

Страница 36: ...of phenol derivatives chlorine compounds or peracetic acid Never use an abrasive Fingerprints on the touchscreen can be cleaned with a dry cloth or some iso propanol Switch off the device at mains sw...

Страница 37: ...perform service work under any circumstances All service work safety inspections repairs may only be performed by service personnel authorized by the manufacturer 9 5 Disposal of the device and access...

Страница 38: ...be switched on No display shown Power supply interrupted Mains cable mains plugs or sockets defect or not properly connected 1 Check that the mains ca ble mains plugs and sockets carry voltage 2 Check...

Страница 39: ...cognitively impaired patients Febrile infections acute infections acute injuries Tuberculosis Haemorrhages or risk of haemorrhage Septic conditions and empyemas Malignant tumors and tumors that are no...

Страница 40: ...e garments may heat up faster and more intensely than the patient s body Synthetic fibers perlon nylon etc are characterized by low absorbency which can cause the skin beneath such fabrics to quickly...

Страница 41: ...were not of a serious nature It is important to note that they are related to the type of therapy not to the device itself Transient aggravation of symptoms Dizziness Erythema Itching Flaking skin 11...

Страница 42: ...105289 58090022 58090019 58090020 58090024 58090023 58090021 For the MicroPro we offer a large variety of accessories Specification Part number Circular field radiator Long field radiator Cradle radia...

Страница 43: ...x 530 mm h w d Weight 44 kg Display 12 1 TFT with capacitive touchscreen Battery CR 2032 Measurement of the rated output power HF power meter Environmental conditions Operation of the de vice Temperat...

Страница 44: ...e radiator Warning The test lamp can be only used to check if the device is emitting microwave power This test doesn t show the level of the emitted microwave power To get an exact inspection of the o...

Страница 45: ...lectronic device For its disposal the according regulations for electronic devices have to be observed On request the manufacturer will provide you with further technical descriptions for all repairab...

Страница 46: ...contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humid ity sh...

Страница 47: ...0 kHz to 80 MHz 6 Veff in ISM and amateur radio bands 3 Veff 150 kHz to 80 MHz 6 Veff in ISM and ama teur radio bands Recommended separation distance d 1 2 P Radiated HF disturbances IEC 61000 4 3 3 V...

Страница 48: ...elements 26 Disposal 37 Duration of treatment 31 E Ending treatment 25 Exclusion of liability 6 F Favorites 33 Focus Radiator 21 Frequency 31 G Grayed out buttons 26 H Help menu 29 I Impulse width 31...

Страница 49: ...card 27 30 Serial number 34 Service menu 34 Side effects 41 Software version 34 Start of operation 22 Switching on 24 Symbols 7 System information 34 T Technical data 43 Therapy 31 Therapy settings 30...

Страница 50: ...50 MicroPro gbo Medizintechnik AG Version 1 4 Notes...

Страница 51: ...MicroPro 51 gbo Medizintechnik AG Version 1 4 Notes...

Отзывы: