Before commencing cleaning and maintenance, the machine must always be
switched off at the main switch and the mains socket.
Checking the impact
dome
During the course of normal use, a deformation / shortening of the rear impact
dome on the applicator heads may occur. As long as the changes are small,
functionality is not affected.
In cases of more extensive deformation of the impact dome or considerable
shortening, the applicator head must be replaced.
To check the proper condition of the applicator head, a test template is
supplied with the device in order to test whether the wear limit has been
reached.
The test should be performed at least once a month, regardless of the
frequency of use.
To do so, the applicator head is removed and measured using the test
template supplied.
Air gap
Applicator OK
Test template
Test template
Template sits flush or there is an air gap at the top
Wear limit reached