10. Backup battery not charging.
a) Blown battery fuse (board mounted F1).
b) Battery not properly plugged into P7
c) Unit not plugged into wall outlet (verify that the
green
AC MAINS indicator is illuminated).
d) Mains AC harness not properly plugged into P1
(verify that the
green
AC MAINS indicator is illuminated).
e) Unit was not permitted to charge for at least 24 hours.
f) Defective battery.
g) Defective battery charging circuitry.
11. AMP FAIL alarm.
a) Transducer(s) amplifier out of range.
b) Battery fully depleted or defective.
c) Extremely high pressure exerted on transducers.
12. Unit cannot be set to STANDBY.
a) Membrane panel not properly plugged into P8.
b) Pressure sensed in the Main or Second cuff
(unit will be alarming “CUFF” “ NOT” “DEFL”).
c) ON/STANDBY not fully pressed.
d) ON/STANDBY button defective.
13. Unit does not turn on.
a) Membrane panel not properly plugged into P8.
b) ON/STANDBY button defective.
c) Blown Fuse(s).
d) Unit not plugged in and battery fully depleted.