Zimmer A.T.S. 1200 Скачать руководство пользователя страница 10

14.

CUFF connector ports

The CUFF connectors are the ports used to connect the
unit to the cuff hoses. Please note that the MAIN CUFF
is the 

red

port and the SECOND CUFF is the 

blue

port.

The 

A.T.S.

1200 Tourniquet is designed and tested for

use with Zimmer single port cuffs. Zimmer does not 
recommend the use of any cuff other than Zimmer single
port cuffs. Do not use dual port cuffs with the 

A.T.S.

1200 Tourniquet.

15.

CUFF indicators

The CUFF indicators illuminate steady when stable cuff
pressure is sensed in the cuff. The CUFF indicators will
also flash during inflation or to indicate an alarm 
condition in the respective cuff.

16.

Pole clamp

The pole clamp is used to mount the unit on an I.V. pole.

NOTE:

Do not hang articles on the tourniquet pole that

are not related to tourniquet use. For stability reasons, do
not use an I.V. pole with a base less than 27.27 inches
(70 cm) in diameter.

2.3 INITIAL SETUP

Inspect to ensure the correct fuse drawer with the 
appropriately rated fuses is present. The 100–120 V unit
uses the 

gray

fuse drawer with 1.0 A time delay fuses. The

220–240 V unit uses the 

black

fuse drawer to 1.0 A time

delay fuses. The power cord should be plugged into the
power entry module on the back of the unit. 

The unit should

be plugged into ~ (AC) power (Mains) for 24 hours before
initial use. During shipping and storage, the unit’s 
battery could become weak. Always charge 24 hours
before any initial use including any calibration checking
procedures, initial checks, tests and any institutional 
performed biomedical evaluations.

2.4 FUNCTIONAL AND CALIBRATION CHECK

The unit shall produce the results explained in the following
steps exactly as indicated. Failure to do so indicates that a
problem may exist and the device is not to be used until 
necessary repair or calibration has been made.

1.

Connect the power plug of the unit to a properly 
polarized and grounded power source with voltage and
frequency characteristics compatible with the 
specifications listed in Section 1.1. Observe that the

green

AC MAINS indicator light turns on.

2.

Turn the unit ON by pressing the ON/STANDBY button
and observe the following:

a)

A “0*0*”/“*0*0” sequence appears on the 
PRESSURE and TIME displays followed by the text
“ATS” “1200” on the respective display.

b)

The unit emits tones when “0*0*”/“*0*0” and text
is displayed.

c)

The unit will display “SELF” “TEST” for an instant.
The unit is self-testing specific system hardware and
software.

d)

“CAL” is displayed in the PRESSURE displays 
during the calibration check.

e)

“0” is displayed in the PRESSURE and TIME 
display after the startup routing is complete. If a
number other than zero is displayed in the 
PRESSURE display, the unit should be calibrated.

3.

Test the PRESSURE set point system as follows:

a)

Press the PRESSURE button.

b)

The PRESSURE display should read “*250” (the
default set point) for 2 seconds.

c)

Within the 2-second time frame, rotate the ROTARY
knob to change the pressure set point (clockwise to
increase, counter-clockwise to decrease). The set
pressure can be maintained between 50 mm Hg and
475 mm Hg in increments of 5 mm Hg.

4.

Test the TIME set point system as follows:

a)

Press the TIME button.

b)

The main TIME display should read “*60” (the
default set point) for 2 seconds.

c)

Within the 2-second time frame, rotate the ROTARY
knob to change the time set point (clockwise to
increase, counter-clockwise to decrease). The set
time can be maintained between 5 and 240 minutes
in increments of 5 minutes.

NOTE:

Anytime an asterisk (*) appears in the left 

display digit, the data being displayed is the set point.
Set pressure and time will revert to the default pressure
and time when the unit is set to STANDBY.

5.

Calibration Check

NOTE:

During the power-up diagnostic self-test

described above, the unit will test calibration. Should an
out of calibration condition be detected, the unit will 
display either “CAL” “FAIL”, “CALM” “FAIL” or
“CAL2” “FAIL” in the PRESSURE and TIME displays.
Even though the unit performs this check at every
power-up, the following quantitative check is 
recommended at regular intervals.

a)

Verify the unit is in the STANDBY mode.

b)

Enter the calibration mode by pressing and holding
the Main cuff Inflate and Deflate buttons while 
powering the unit ON. The unit will enter the 
calibration mode when the Pressure and Time 
display momentarily displays “CAL” “MODE”.
After the “CAL” “MODE” is displayed, the unit
will display the software revision level. The 
software revision level can be recorded for future
reference.

NOTE:

The calibration is only being checked in this section.

For complete calibration, see Maintenance Section 3.0.

8

A.T.S. 1200 TOURNIQUET SYSTEM

Содержание A.T.S. 1200

Страница 1: ...Operator Service Manual A T S 1200 TOURNIQUET SYSTEM REF 60 1200 101 00...

Страница 2: ...Inc be liable for any special incidental or consequential damages whether based on breach of warranty or other legal theory whether or not such damages are foreseeable Some states do not allow limita...

Страница 3: ...ibration Check 8 2 5 Pressure and Time Defaults 9 2 6 Single Cuff Operation 10 2 7 Dual Cuff Operation 11 2 8 Bier Block Cuff Operation IVRA 11 2 9 Alarm Conditions 11 3 0 MAINTENANCE 3 1 General Main...

Страница 4: ...erview front 23 2 A T S 1200 Tourniquet Overview rear 24 3 Calibration Setup Overview 25 4 Disassembly front case opening 26 5 Disassembly front case removal 26 6 Disassembly disconnecting wiring 26 7...

Страница 5: ...pressure set point Can also be pressed to verify the set point TIME Button Used in conjunction with the ROTARY knob to adjust the time alarm set point Can also be pressed to verify the set point MAIN...

Страница 6: ...nt light emitting diodes LED Displays pressure setting sensed cuff pressure and hardware failure conditions other messages TIME Display Red 14 segment light emitting diodes LED Displays time alarm set...

Страница 7: ...ll cause bacteria exotoxins or malignant cells to spread to the general circulation or where it could dislodge thromboemboli that may have formed in the vessels The tourniquet cuff must be applied in...

Страница 8: ...body level Whenever IVRA Bier Block anesthesia is used it is recommended that the tourniquet remain inflated for at least 20 minutes from the time of injection WARNING Cuffs will not deflate in STANDB...

Страница 9: ...ely 2 seconds before the unit will allow a cuff to deflate 7 SECOND CUFF INFLATE button Inflation of the Second cuff is initiated by depressing the blue INFLATE button 8 SECOND CUFF DEFLATE button Def...

Страница 10: ...ys followed by the text ATS 1200 on the respective display b The unit emits tones when 0 0 0 0 and text is displayed c The unit will display SELF TEST for an instant The unit is self testing specific...

Страница 11: ...e the pressure to 250 mmHg The PRESSURE display should read 250 5 mmHg n Increase the pressure to 475 mmHg The PRESSURE display should read 475 5 mmHg o Remove the calibration hose from the unit The P...

Страница 12: ...ressure during the procedure 4 Prepare the patient in accordance with your established procedures and cuff manufacturer s instructions The precautions of Section 1 and the following are offered as a g...

Страница 13: ...re is a significant leak in the original cuff this feature could cause the inflation rate of the subsequent cuff to be longer and perhaps even cause the 30 second inflation alarm to sound The display...

Страница 14: ...ude that a hardware failure has occurred rendering the unit unusable The appropriate action is to set the unit to STANDBY by pressing the ON STANDBY button Since this removes power from the internal i...

Страница 15: ...the set value or greater at power up to determine if it needs resetting CUFF NOT DEFLATED normal CUFF Check for kinks in hose If alarm persists Pressure in deflated cuff is NOT disconnect hose from c...

Страница 16: ...ice the unit combination occurred WATCHDOG FAILURE WDT FAIL Cycle the ON STANDBY Button If problem Windowing watchdog system persists service the unit detected a malfunction SYSTEM FAILURE SYS FAIL Cy...

Страница 17: ...very six months 1 Cleaning The exterior of the unit may be cleaned with a cloth that has been dampened not dripping with a mild detergent The interior of the unit may be vacuumed or blown out as requi...

Страница 18: ...50 mm Hg to the unit s Main cuff port Once the pressure has stabilized press the PRESSURE button so the unit can calibrate the 250 mm Hg point The unit will beep to let the user know the set point was...

Страница 19: ...steps in Section 2 4 Step 5 Calibration Check before using this unit on a patient 3 5 LEAK TESTING The A T S 1200 Tourniquet is capable of keeping a cuff with a substantial leak inflated Naturally it...

Страница 20: ...ilure that may occur is when an audible alarm occurs that cannot be silenced by the ALARM SILENCE button The valve and pump will be disabled which seals off the cuff to prevent pressure loss The displ...

Страница 21: ...not properly plugged into P8 f Defective AC MAINS indicator g Defective AC MAINS indicator circuitry 4 No flashing orange BATTERY indicator light a Unit running on AC b Membrane Panel not properly plu...

Страница 22: ...ve battery charging circuitry 11 AMP FAIL alarm a Transducer s amplifier out of range b Battery fully depleted or defective c Extremely high pressure exerted on transducers 12 Unit cannot be set to ST...

Страница 23: ...ry use only TP19 0 VDC N A Digital ground TP20 N A N A Factory use only TP21 26 4 to 27 5 VDC 1 0 VDC Boost supply for battery charger TP22 5 VDC 0 5 VDC Supply voltage for battery charger TP23 5 VDC...

Страница 24: ...263 Calibration Hose 0600 2002018 Plumbing Assembly 60 0938 003 00 A T S Stand 5 Wheel Caster Kit 60 2360 001 00 A T S Pole Clamp Knob w Screw 60 3000 001 00 A T S Cal Regulator Kit 60 7000 027 00 Bac...

Страница 25: ...ivation Button G ON STANDBY Button H Green AC MAINS Indicator I Orange BACKUP BATTERY Indicator J SECOND CUFF INFLATE DEFLATE Buttons K Second Cuff Port L Main Cuff Port M MAIN CUFF INFLATE DEFLATE Bu...

Страница 26: ...R Power Cord S Cord Retaining Clip T Factory Test Port U Pole Clamp V Mains Fuse Block 24 2...

Страница 27: ...W Calibrated Pressure Meter with a range of 0 to 500 mm Hg X Pressure Regulator Source adjustable from 0 to 500 mm Hg Y Calibration Hose Included with Unit 25 3...

Страница 28: ...the front case 1 Carefully slide the front off the rear Disconnecting the wiring 1 Use caution when disconnecting the wiring Mark the wires before disconnecting or see Section 3 2 for proper re connec...

Страница 29: ...Rear Case separation 1 All components are easily accessed when the rear case is removed Reverse the process to reassemble Be extremely careful not to pinch any wiring or tubing when reassembling 8 27...

Страница 30: ...28 9 Control Board Layout...

Страница 31: ...Equipment with respect to electric shock fire and mechanical hazards only in accordance with UL 60601 1 CAN CSA C22 2 No 601 1 and IEC 60601 1 This product contains electrical or electronic materials...

Страница 32: ...30 10...

Страница 33: ...SCHEMATICS...

Страница 34: ...Revised 04 06 Zimmer Orthopaedic Surgical Products 2002 2006 Zimmer Orthopaedic Surgical Products Inc 200 West Ohio Avenue Printed in U S A P O Box 10 62 8000 015 00 ENG Dover Ohio 44622 U S A 0123...

Отзывы: