DCO-1318
Effective Date: 01-SEP-2021
LBL 6028 Rev. G
Page 19 of 23
Maximum RF Voltage Output
Mode
Maximum Voltage
Cut
450 V
Coag
200 V
III.
Output Power vs. Setting
Setting
Power (W)
1
195
2
215
3
235
4
255
5
275
6
300
7
320
8
345
9
370
10
400
COAG 1
55
COAG 2
75
COAG 3
95
APPENDIX G EMC TABLES
The following tables provide information on the electromagnetic environment in which the Tricera System is capable of operating
safely. Use of accessories and cables other than those specified may result in increased emissions or decreased immunity of the
system. To ensure proper grounding reliability, a Hospital Grade Power Cord must be used with a receptacle marked “Hospital
Grade”.
List of Tricera Controller Accessories:
•
Tricera Fluid Management Accessories
•
Tricera Footswitch
•
10 ft. Hospital Grade Power Cord
Table 1: Electromagnetic Emissions Statement
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Tricera System is intended for use in the electromagnetic environment specified below. The customer or the user of the Tricera System should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment
–
Guidance
RF emissions CISPR 11
Group 1
The Tricera System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions CISPR 11
Class A
The Tricera System is suitable for use in all establishments other
than domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies