DCO-1318
Effective Date: 01-SEP-2021
LBL 6028 Rev. G
Page 17 of 23
Problem
Possible Causes
Remedy
Display Message
Purge stops before
complete
-
Inflow tube is obstructed
during purge
-
Scope/sheath is inserted into
joint during purge
Check inflow tube and
scope/sheath
connections
Remove scope/sheath
from patient
F10: Pressure Unstable
Retry Purge
F11: Scope/Cannula in Joint.
Please Remove and Retry
Purge
Cannot ACTIVATE
or COAG
-
Handpiece does not Activate
or COAG
Ensure that normal
saline [sodium chloride
(0.9% w/v; 150mmol/l)] is
being used as irrigation
solution
Ensure that the
footswitch is plugged into
the gray port on the
Controller
Ensure the Handpiece is
securely plugged into the
connector
Replace Tricera Device
N/A
N/A
N/A
F12: Handpiece Cannot
Position Window Properly.
Please Retry
Insufficient Suction
-
Suction Tubing not
connected correctly
-
Suction tubing kinked or
occluded
-
Tricera Device defective
Check that Suction
Tubing is properly
inserted in pump, check
that connections are
secure, replace if
necessary
Check Suction tubing for
occlusion
Replace Shaver blade
N/A
Handpiece does not
power on
-
Handpiece is defective
-
Controller is defective
Ensure the Handpiece is
securely plugged into the
connector
Replace Controller
Replace Handpiece
Send Controller and
Handpiece in for repair
OLED display on Handpiece
does not turn on.