15
REPLACING THE BATTERIES
WARNING
: To avoid electric shock, disconnect the test
leads from any source of voltage before removing the
battery door.
1. When the batteries become exhausted or drop below
the operating voltage, “BAT” will appear in the
right-hand side of the LCD display. The batteries should
be replaced.
2. Follow instructions for installing batteries. See the
Battery Installation section of this manual.
3. Dispose of the old batteries properly.
WARNING:
To avoid electric shock, do not operate your
meter until the battery door is in place and fastened
securely.
REPLACING THE FUSES
WARNING
: To avoid electric shock, disconnect the test
leads from any source of voltage before removing the
fuse door.
1.
Disconnect the test leads from the meter and any
item under test.
2.
Open the fuse door by loosening the screw on the
door using a Phillips head screwdriver.
3.
Remove the old fuse from its holder by gently pulling
it out.
4.
Install the new fuse into the holder.
5.
Always use a fuse of the proper size and value