4. Care and Maintenance  

  

  

– Method of Light Source Disinfecting 

 

  

  

WARNING:  

Do not use: Ethylene Oxide, autoclave, ultrasonic, chlorides, formaldehyde, or 

hydrogen peroxide.

  

Using any of these may damage the light source and will void the product 

warranty.   
  

The HLS is sealed for protection against ingress of fluids only when assembled. The entire light source is 

submersible in water and disinfectant solutions. However, the light source should not remain immersed in 
water or cleaning solutions for prolonged periods of time. After the disinfection cycle has concluded, remove 
the light source from the disinfection solution, dry and store it in a clean and dry environment.  

  
A thorough cleaning and rinsing are the first and most important steps in disinfecting this light source. Without 

a thorough cleaning and rinsing, it may not be possible to achieve high-level disinfection. Cleaning is the 
removal of all adherent visible soil (i.e. blood, protein substances and/or other debris) from the surface, 
crevices and joints of the light source.  

  
Use caution when cleaning and disinfecting as some methods may be harmful to the light source and could 

result in extensive damage. Manual cleaning is the recommended method for cleaning the light source. To be 
effective, cleaning agents must assist in the removal of residual organic debris without damaging the device. 
No single cleaning agent removes all types of debris or is safe to use with every type of reusable device. 
Certain cleaning agents may damage the device’s materials. Cleaning agents should be used in the correct 
dilution/concentration and at the correct temperature in accordance with the cleaning agent manufacturer’s 
directions. Please contact Zibra Corp at 1-800-758-8773 to verify the compatibility of a cleaning method not 
listed below under Section II - 

Light Source Disinfecting Protocol

.  

  

  

  

WARNING: 

Remove battery prior to disinfecting.

 

Ensure Power Module is fully inserted on the 

Heatsink, so that the o-ring seal will prevent liquid intrusion into the battery compartment.

  

 

   
  

WARNING

: With battery removed, rotate Intensity Control Knob until it clicks into the maximum 

brightness setting of 10, thereby preventing accidental separation while handling. 

 

 

  

 

II - Light Source Disinfecting Protocol 
 

 

1.  Pre-Cleaning with detergent 
2.  Enzymatic Cleaning and Rinsing; Enzol Fast Acting Enzymatic Cleaner or equivalent  
3.  High-Level Disinfection with CIDEX OPA (0.55% ortho-Phthalaldehyde Solution)   

  

Pre-Cleaning 

 

  

Zibra recommends the light source be pre-cleaned immediately after the device has become contaminated. 

This will prevent patient material from adhering to the light source. Dried material is difficult to remove and 
can render the disinfection process ineffective. Gently wipe all debris from the exterior surfaces using a soft, 
lint-free cloth and water to which you have added a low sudsing detergent, diluted in accordance with the 
detergent manufacturer’s instructions.  
  
 

 

Содержание ZC-LS006-1002

Страница 1: ...is covered under US Patents issued and pending Handheld LED Light Source HLS Operator s Manual Sales and Service by Zibra Corp 640 American Legion Hwy Westport MA 02790 USA USA Tel 1 800 758 8773 www...

Страница 2: ...2 This page intentionally left blank...

Страница 3: ...mended Separation Distance between Portable and Mobile RF Communications Equipment 14 7 Repair Service 14 8 Warranty 14 9 Symbol Descriptions 14 WARNING This device is not supplied sterile Clean and d...

Страница 4: ...oy this light source such as endoscopy WARNING Avoid direct viewing of the high intensity light at the front of the Handheld Light Source Never point it s light at other s This manual describes the pr...

Страница 5: ...the Off position 0 will allow the Power Module to be pulled out An o ring seals the battery and electronics when the Power Module is inserted into the Heatsink Handle On Off Switch The On Off switch t...

Страница 6: ...Operating Storage and Transport Relative Humidity Range 0 95 RH Battery Recharge Cycles 500 recharges RoHs Compliant Yes Accessories Light Sources Accessories Listing Part Number Handheld Light Source...

Страница 7: ...is finger tight Do not over tighten ACMI or Wolf Insert the light guide post of the endoscope into the Scope Adapter until it snaps into place It is normal for a small amount of play to exist between...

Страница 8: ...Power Module by carefully aligning the key on the bottom of the Battery Sled with the keyway cut in the Heatsink Handle NOTE The key must be aligned with the keyway before full insertion is possible P...

Страница 9: ...es the battery is fully charged or no battery is present 3 After the battery is fully charged pull the battery s out of the charger 4 Unplug the AC adapter when not in use to conserve energy CAUTION N...

Страница 10: ...le cleaning agent removes all types of debris or is safe to use with every type of reusable device Certain cleaning agents may damage the device s materials Cleaning agents should be used in the corre...

Страница 11: ...t as a result of rinsing with potable water 6 Dry the exterior of the light source with a soft clean towel III Drying the Light Source After the HLS becomes wet either from water or cleaning fluids th...

Страница 12: ...leared off Dry with compressed air or gas Light source becomes hot after prolonged use at full power Light source has been left on for more than 30 minutes at full power Turn the light source to a low...

Страница 13: ...n portable and mobile RF Communications Equipment and the HLS according to the maximum output power of the communications equipment Note The HLS is internally battery powered and has no I Os Conducted...

Страница 14: ...ny person not authorized by Zibra will result in immediate loss of warranty THE ABOVE WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES EITHER EXPRESS OR IMPLIED INCLUDING ANY WARRANTY OF MERCHANTABILITY...

Страница 15: ...____ _______________________________________ _______________________________________ _______________________________________ _______________________________________ ___________________________________...

Страница 16: ...16 Part No LS006 0057 Aug 2018...

Отзывы: