①
Install the GND15 so that the prism surface is at a
right angle to the ground.
②
Attach the flow pipe to the GND15 with
O-ring (accessory) inserted between them, and fasten
them together with the clamp band (accessory).
③
The prism surface may become contaminated with
solids, dirt and/or grease. If this happens, the prism
surface must be cleaned by hand.
④
2 PU pipe connect with inlet and output separately
Note :
Cleaning the Prism instruction in the next page
The sample inlet unit should be installed in such a
manner that it can be easily removed to allow access
to the prism for cleaning
Note :
Suspending the GND15 (2.5kg) when using the
sample inlet unit to connect to the piping is dangerous ,
Use of the optional stand is recommended.
ZHYQ Sensor&Control Page 11 of 15
Mounting procedure:
In-line Refractometer
Model : GND15
(
high temperature Application
)