GND15(B)  Intrinsically  safe  In-line  Refractometer,  which  is  composed  of  light  source,  sapphire  refraction,  light 
receiver and signal modulation circuit. The light source emits a stable light signal through the sapphire to contact 
the  measured  medium,  and  the  concentration  of  the  measured  medium  produces  a  corresponding  signal.  The 
optical receiver calculates the refractive index of the collected refracted light source to obtain the corresponding 
concentration  through  the  data  algorithm.  Since  the  concentration  will  be  affected  by  temperature,  internal 
temperature calibration is adopted to reduce the influence of temperature drift.

The measuring principle diagram as below:

Automatic temperature compensation

Sturdy and durable, compact and easy to install

Easy to operate, easy to connect with other devices for automatic control

Inline real-time monitoring and control of liquid substance concentration in various industries

The  GND15(B)  intrinsically  safe  concentration  meter  is  mainly  used  for  the  preparation  and  detection  of  the 

weight percentage concentration of the cooling and cooling emulsion in the machining industry and the M-10 and 

MDT  emulsions  commonly  used  in  mines.  It  provides  convenience  for  safe  and  economical  preparation  of 

emulsions.  The  method  is  widely  used  in  mechanical  processing,  mining  selection,  underground  support  and 

other  fields.  It  is  used  to  detect  the  weight  percentage  concentration  of  water-soluble  mine  emulsion,  which 

provides great convenience for accurately preparing emulsion-pressed liquefied liquid.

Working Principle

Applications

Features