1. When there is no signal output from the concentration meter, firstly check the disconnection or short circuit of 
the extension cable or whether the working voltage of the concentration meter is normal

2. If the concentration meter is suspected to be malfunctioning, the following methods can be used

Remove the densitometer, measure water and a certain concentration of medium respectively, measure with the 
mA current of the multimeter, blue electronic wire: 4~20 output positive and yellow electronic wire: 4~20 output 
negative, the current signal should be 4mA, and the measured The current output corresponding to the medium 
is normal, otherwise the concentration meter is faulty.

Periodic/timed maintenance (operable by cleaning unit)
PS:  we  can  also  supply  Auto  or  manual  washing  device  for  optional,  if  you  need,  Please  let  us  know 
when order. 

Wipe the glass of the densitometer probe clean, add 0.3mL of tap water, if the test value is not equal to 4mA, 
short-circuit the zero line to clear it.

Concentration  meters  measuring  high  temperature,  corrosive  media  or  operating  under  vibration  conditions 
should be calibrated once every six months.

After the concentration meter has been running for 5 years, an insulation resistance test should be performed, 
and the insulation resistance should be above 

500MΩ.

Concentration value vs. Refractive Index graph

Normal troubleshooting

Maintenance