CAUTION

* The patient can measure ,transmit 

 

data and change batteries under normal circumstances and maintain 

the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated 

interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient environment. If you are 

allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and 

found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any 

potential sensation or irritation reaction.

* If you experience discomfort during a measurement, such as pain in the wrist or other complaints, press 

the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from 

your wrist.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not 

deflate when pressure reaches 40 kPa (300 mmHg), detach the cuff from the wrist and press the 

START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. Check the device, 

do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury, 

improper results, or serious danger.

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The 

typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after maintenance and 

repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air 

leakage (testing at least at 50mmHg and 200mmHg).

*  Manufacturer will make available on request circuit diagrams, component part lists, descriptions, 

calibration instructions,etc., to assist to service personnel in parts repair.

 * The operator shall not touch output of batteries and the patient simultaneously.

*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff 

immediately. Prolonged high pressure (cuff pressure 

 300mmHg or constant pressure 

 15mmHg for 

more than 3 minutes) applied to the wrist may lead to an ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of patient blood 

circulation.

* When measurement, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically 

investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.

* This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing 

inaccurate readings, the effects of this device on the fetus are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, 

the patient’s wrist and fingers will become anaesthetic, swollen and even purple due to a lack of blood.

* When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust 

and direct sunlight. Never place any heavy objects on the storage case.

* This device may be used only for the purpose described in this booklet. The manufacturer cannot be held 

liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. Observe the storage and 

operating conditions described in this booklet.

* The maximum temperature that the applied part can be achieved is 42.5

 while the environmental 

temperature is 40

.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable 

anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

CAUTION

INTRODUCTION

INTRODUCTION

7

6

Содержание WS-380

Страница 1: ...d Pressure Monitor WS 380 Please keep this manual well in order to reference in future 2019 07 16 70G 85 70 mm TMB 1598 BS 006 2019 07 16 2019 07 16 2019 07 16 TMB 1598 BS 006 GB 09 PO 2F4198 TMB 1598...

Страница 2: ...31 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbeat detector Why does my blood pressure fluctuate throughout the day Why do I g...

Страница 3: ...rement pulse rate measurement and the result storage Readings taken by the WS 380 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method This manual...

Страница 4: ...The device is not intended for patient transport outside a healthcare facility The device is not intended for public use This device is intended for no invasive measuring and monitoring of arterial bl...

Страница 5: ...el in parts repair The operator shall not touch output of batteries and the patient simultaneously On the rare occasion of a fault causing the cuff to remain fully inflated during measurement open the...

Страница 6: ...to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS Wireless communications equipment such as wireless home network devices mobile phones cor...

Страница 7: ...d save the results for user 1 User 2 Start measurement and save the results for user 2 Blood pressure monitor is detecting a heartbeat during measurement Blood pressure monitor is detecting an irregul...

Страница 8: ...e 5 Cuff List 1 Blood Pressure Monitor WS 380 2 2 AAA batteries 3 User manual Installing and Replacing the Batteries Slide off the battery cover Install the batteries by matching the correct polarity...

Страница 9: ...to the environment Do not dispose with daily garbage Remove the old batteries from the device following your local recycling guidelines Setting Date Time Positioning Symbol and Measurement Unit 1 It...

Страница 10: ...s of setting it will turn off within 1 minute 2 Press button to change the YEAR Each press will increase the numeral by one in a cycling manner 3 When you get the right year press button to confirm an...

Страница 11: ...then the monitor will turn off Before you start the measurement please select the desired user ID first When the blood pressure monitor is off press and hold button until the user ID blinks 1 Select...

Страница 12: ...the cuff should be at the level of the right atrium of the heart Before starting measurement please sit comfortably with legs uncrossed feet flat on the floor back and wrist supported Rest for 5 minut...

Страница 13: ...h data stored in your mobile device 2 If the data transmission fails the symbol will remain The pending measurement data will be transmitted to your mobile device when next measurement is complete Tip...

Страница 14: ...end should be reasonably close from 1 meter to 10 meters Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range To avoid interfere...

Страница 15: ...n the desired user ID blinks press button to confirm Then press button to check the records of the selected user 3 1 2 If you did not get the correct measurement you can delete all the results for the...

Страница 16: ...nner or drinking Within 20 minutes after taking a bath Measurements may be inaccurate if taken in the following circumstances Immediate measurement after tea coffee smoking When talking or moving your...

Страница 17: ...ic pressure What is the standard blood pressure classification The chart on the right is the standard blood pressure classifi cation published by American Heart Association AHA Please consult a physic...

Страница 18: ...ties at an early stage Why does my blood pressure fluctuate throughout the day 1 Individual blood pressure varies multiple times everyday It is also affected by the way you tie your cuff and your meas...

Страница 19: ...ing on the arm and then measure again The monitor detected motion while measuring E 04 shows Relax for a moment andthen measure again E 12 shows PROBLEM SYMPTOM CHECK THIS REMEDY Error message EExx sh...

Страница 20: ...a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of 700 hPa to 1060 hPa Temperature 20 C to 60 C A relative humidity range of 93 non condensing at a water vapour pres...

Страница 21: ...ll not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is enc...

Страница 22: ...l requirements EN 1060 3 1997 A2 2009 Non invasive sphygmomanometers Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems IEC 80601 2 30 2009 A1 2013 Medical elect...

Страница 23: ...of this equipment and result in improper operation Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm...

Страница 24: ...urst IEC 61000 4 4 Surge IEC61000 4 5 IEC 60601 1 2 Test level 2 kV 4kV 8 kV 15 kV air Not application Not application Not application Not application Not application Not application 30 A m 50Hz 60Hz...

Страница 25: ...GMRS 460 FRS 460 FM c 5kHz deviation 1kHz sine 2 0 3 28 710 704 787 745 780 LTE Band 13 17 Pulse modulation b 217Hz 0 2 0 3 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pu...

Отзывы: