background image

Manufactured for Zewa, Inc.

12960 Commerce Lakes Drive#29

Fort Myers, FL 33913 USA

www.zewa.com

Toll Free Customer Service:

1-888-993-3592

[email protected] 

version:1.0

User Manual

Blood Pressure Monitor

UAM-880

To use the monitor correctly and safely, please read the manual thoroughly.

Thank you  for purchasing a ZEWA Blood Pressure Monitor .

Arm Type

Содержание UAM-880

Страница 1: ...913 USA www zewa com Toll Free Customer Service 1 888 993 3592 warranty zewa com version 1 0 User Manual Blood Pressure Monitor UAM 880 To use the monitor correctly and safely please read the manual thoroughly Thank you for purchasing a ZEWA Blood Pressure Monitor Arm Type ...

Страница 2: ...ords Delete the Records INFORMATION FOR USER 20 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 22 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbeat detector Why does my blood pressure fluctuate throughout the day Why do I get a different blood pressure at home compared to the hospital Is the result the same if measuring...

Страница 3: ...the license symbol of a European network of industry funded systems for recycling the packaging materials of consumer goods Symbol for RECYCLE Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF APPLIED PARTS Symbol for DIRECT CURRENT Symbol for ENVIRONMENT PROTECTION Electrical waste products should not be disposed of with household wast...

Страница 4: ... When measurement please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060 2 2013 To verify the calibration of the AUTOMATED SPHYGMOMANOMETER please contact the manufacturer Th...

Страница 5: ...gth At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCU...

Страница 6: ...tteries whenever the below happens The shows The display dims The display does not light up In order to get the best effect and protect your monitor please use the the right batteries and special power adapter which complies with local safety standard CAUTION CAUTION Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Bat...

Страница 7: ...d pressing button for 3 seconds to enter the year setting mode 2 Press the button to change the YEAR 3 When you get the right year press button to confirm and turn to next step 4 Repeat step 2 and 3 to set the MONTH and DAY 5 Repeat steps 2 and 3 to set the HOUR and MINUTE 6 After the MINUTE is set the LCD will display donE first then display all the settings you have done and then turn off 11 10 ...

Страница 8: ...u with poor circulation in your left arm use your right arm Roll or push up your sleeve to expose the skin Make sure your sleeve is not too tight 2 3 5 Sit comfortably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first m...

Страница 9: ... Start the Measurement 4 Press button to power off otherwise it will turn off within 1 minute 2 This device will proceed to data transmission after measurement The Bluetooth symbol blinks on the LCD indicates data is transmitting 3 If the data is successfully transmitted the Bluetooth symbol will light up and the will disappear If the data transmission fails the device will turn off 15 14 DATA MAN...

Страница 10: ...m 1 meter to 10 meters Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range To avoid interference other electronic devices particularly those with wireless transmission Transmitter should be kept at least 1 meter away from the monitor Recall the Records 2 Press the button again to rotate the records The order of the record dat...

Страница 11: ... user by following steps below Delete the Records 4 If there is no record the below display will be shown 2 Press button to confirm deleting all the memories the LCD will display User ID dEL donE and the monitor will turn off 3 To exit the delete mode without deleting any records press button before pressing button to confirm any delete commands 19 18 DATA MANAGEMENT DATA MANAGEMENT ...

Страница 12: ... urine Measurements may be inaccurate if taken in the following circumstances Maintenance In order to get the best performance please follow the instructions below Put in a dry place and avoid the sunshine Avoid intense shaking and collisions Using wet cloths to remove dirt Avoid touching water clean it with a dry cloth in case Avoid washing the cuff Avoid dusty and unstable temperature environmen...

Страница 13: ...ry time What you need to pay attention to when you measure your blood pressure at home If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the upper arm If you feel anxious Taking 2 3 deep breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down What is the standard blood pressure classification The chart ...

Страница 14: ...5 C to 40 C A relative humidity range of 15 to 90 non condensing but not requiring a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of 700 hPa to 1060 hPa Temperature 20 C to 60 C A relative humidity range of 93 non condensing at a water vapour pressure up to 50hPa PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Batt...

Страница 15: ...here is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between ...

Страница 16: ...ts for basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment EN 60601 1 2 2015 IEC 60601 1 2 2014 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic disturbances Requirements and tests EN ISO...

Страница 17: ...ng of the transmitter in watts W according to the transmitter manufacturer d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol P P P P a b Guidance and manufa...

Страница 18: ...NUFACTURER should consider reducing the minimum separation distance based on RISK MANAGEMENT and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation E Where P is the maximum power in W d is the minimum separation distance in m and E is the IMM...

Отзывы: