background image

 

EMC Guidance

1)This product needs special precautions regarding EMC and needs to be 

installed and put into service according to the EMC information provided,

and this unit can be affected by portable and mobile RF communications 

equipment.

2)* Do not use a mobile phone or other devices that emit electromagnetic 

fields, near the unit. This may result in incorrect operation of the unit.

3)Caution: This unit has been thoroughly tested and inspected to assure 

proper performance and operation!

4)* Caution: This machine should not be used adjacent to or stacked with 

other equipment and that if adjacent or stacked use is necessary, this 

machine should be observed to verify normal operation in the configuration in 

which it will be used.

Table 1 

Guidance and manufacturer’s declaration – electromagnetic emissions

RF emissions 
CISPR 11

Group 1 

Class B 

Class A

Complies

Compliance

Harmonic emissions 
IEC 61000-3-2

Voltage fluctuations/ 
flicker emissions  
IEC 61000-3-3  

RF emissions 
CISPR 11

Emissions test

Electromagnetic environment - guidance

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

The device uses RF energy only for its internal 

function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

The device is suitable for use in all establishments,

other than domestic and those directly connected to 

the public low-voltage power supply network that 

supplies buildings used for domestic purposes.

Complied Standards List

EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Application 

of risk management to medical devices

EN 1041:2008 Information supplied by the manufacturer of medical 

devices

EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical 

electrical equipment - Part 1: General requirements for basic safety and 

essential performance

EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical 

equipment - Part 1-11: General requirements for basic safety and 

essential performance - Collateral standard: Requirements for medical 

electrical equipment and medical electrical systems used in the home 

healthcare environment

EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical 

equipment - Part 1-2: General requirements for basic safety and 

essential performance - Collateral standard: Electromagnetic 

disturbances - Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: 

Requirements and test methods for non-automated measurement type

EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - Part 3: 

Supplementary requirements for electro-mechanical blood pressure 

measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test 

procedures to determine the overall system accuracy of automated 

non-invasive sphygmomanometers

EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 Medical 

electrical equipment - Part 1-6: General requirements for basic safety 

and essential performance - Collateral standard: Usability

IEC 62366-1:2015 Medical devices - Part 1: Application of usability 

engineering to medical devices

EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015   Medical 

device software - Software life-cycle processes

Risk management

Labeling

User manual

General Requirements 

for Safety

Electromagnetic

compatibility

Performance

requirements

Clinical investigation

Usability

Software life-cycle 

processes

Bio-compatibility

ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: 

Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: 

Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: 

Tests for irritation and skin sensitization

EN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. 

Symbols to be used with medical device labels, labelling and 

information to be supplied. Part 1 : General requirements

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 

2-30: Particular requirements for the basic safety and essential 

performance of automated non-invasive sphygmomanometers

ISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: 

Clinical validation of automated measurement type

28

29

EMC GUIDANCE

EMC GUIDANCE

Содержание UAM-880

Страница 1: ...913 USA www zewa com Toll Free Customer Service 1 888 993 3592 warranty zewa com version 1 0 User Manual Blood Pressure Monitor UAM 880 To use the monitor correctly and safely please read the manual thoroughly Thank you for purchasing a ZEWA Blood Pressure Monitor Arm Type ...

Страница 2: ...ords Delete the Records INFORMATION FOR USER 20 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 22 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbeat detector Why does my blood pressure fluctuate throughout the day Why do I get a different blood pressure at home compared to the hospital Is the result the same if measuring...

Страница 3: ...the license symbol of a European network of industry funded systems for recycling the packaging materials of consumer goods Symbol for RECYCLE Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF APPLIED PARTS Symbol for DIRECT CURRENT Symbol for ENVIRONMENT PROTECTION Electrical waste products should not be disposed of with household wast...

Страница 4: ... When measurement please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060 2 2013 To verify the calibration of the AUTOMATED SPHYGMOMANOMETER please contact the manufacturer Th...

Страница 5: ...gth At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCU...

Страница 6: ...tteries whenever the below happens The shows The display dims The display does not light up In order to get the best effect and protect your monitor please use the the right batteries and special power adapter which complies with local safety standard CAUTION CAUTION Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Bat...

Страница 7: ...d pressing button for 3 seconds to enter the year setting mode 2 Press the button to change the YEAR 3 When you get the right year press button to confirm and turn to next step 4 Repeat step 2 and 3 to set the MONTH and DAY 5 Repeat steps 2 and 3 to set the HOUR and MINUTE 6 After the MINUTE is set the LCD will display donE first then display all the settings you have done and then turn off 11 10 ...

Страница 8: ...u with poor circulation in your left arm use your right arm Roll or push up your sleeve to expose the skin Make sure your sleeve is not too tight 2 3 5 Sit comfortably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first m...

Страница 9: ... Start the Measurement 4 Press button to power off otherwise it will turn off within 1 minute 2 This device will proceed to data transmission after measurement The Bluetooth symbol blinks on the LCD indicates data is transmitting 3 If the data is successfully transmitted the Bluetooth symbol will light up and the will disappear If the data transmission fails the device will turn off 15 14 DATA MAN...

Страница 10: ...m 1 meter to 10 meters Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range To avoid interference other electronic devices particularly those with wireless transmission Transmitter should be kept at least 1 meter away from the monitor Recall the Records 2 Press the button again to rotate the records The order of the record dat...

Страница 11: ... user by following steps below Delete the Records 4 If there is no record the below display will be shown 2 Press button to confirm deleting all the memories the LCD will display User ID dEL donE and the monitor will turn off 3 To exit the delete mode without deleting any records press button before pressing button to confirm any delete commands 19 18 DATA MANAGEMENT DATA MANAGEMENT ...

Страница 12: ... urine Measurements may be inaccurate if taken in the following circumstances Maintenance In order to get the best performance please follow the instructions below Put in a dry place and avoid the sunshine Avoid intense shaking and collisions Using wet cloths to remove dirt Avoid touching water clean it with a dry cloth in case Avoid washing the cuff Avoid dusty and unstable temperature environmen...

Страница 13: ...ry time What you need to pay attention to when you measure your blood pressure at home If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the upper arm If you feel anxious Taking 2 3 deep breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down What is the standard blood pressure classification The chart ...

Страница 14: ...5 C to 40 C A relative humidity range of 15 to 90 non condensing but not requiring a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of 700 hPa to 1060 hPa Temperature 20 C to 60 C A relative humidity range of 93 non condensing at a water vapour pressure up to 50hPa PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Batt...

Страница 15: ...here is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between ...

Страница 16: ...ts for basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment EN 60601 1 2 2015 IEC 60601 1 2 2014 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic disturbances Requirements and tests EN ISO...

Страница 17: ...ng of the transmitter in watts W according to the transmitter manufacturer d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol P P P P a b Guidance and manufa...

Страница 18: ...NUFACTURER should consider reducing the minimum separation distance based on RISK MANAGEMENT and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation E Where P is the maximum power in W d is the minimum separation distance in m and E is the IMM...

Отзывы: