5. Cleaning and Disinfection Method
After each use, it is necessary to clean and disinfect the cup components (including spray nozzle,
medication cover), mask and mouthpiece. The specific methods of cleaning and disinfection are
recommended as follows:
1. Cleaning
Please turn o� the power when cleaning the device. Do not connect the device with power supply.
1) Remove the components from the main unit: the medication cup components (including the spray
nozzle, medication cover), the inhalation mask or the mouthpiece, and soak them in clean warm
water (which is no more than 40 degrees centigrade) for about 5 minutes.
2) After cleaning, wipe all the components with clean and sterile medical gauze, and let them dry
su�ciently.
3) Wipe the outer shell of main unit. If there is medication residue at the electrode contact, please
clean it with wet sterile medical gauze. After cleaning, keep the main unit dry.
4) Store all the parts in a dry and clean place to avoid contamination.
10
2. Disinfection
After each use, it is necessary to disinfect the components of the cup, including spray nozzle,
medication cover, mask, mouthpiece, etc., as follows:
1)Disinfection with hydrogen peroxide
Disinfect all the components by placing them in 2% hydrogen peroxide for 10 minutes, including
medication cup components (such as spray nozzle and medication cover), mask and mouthpiece,
or wipe and disinfect the medication cup components (including spray nozzle, medication cover),
mask and mouthpiece. After disinfection, rinse all the parts with distilled water, then wipe with clean
and sterile medical gauze or air-dry naturally to keep all parts dry.
Notes: The main unit cannot be washed with water to prevent water from entering the main unit.
Use clean sterile gauze to wipe the moisture on the main unit and components and keep them
dry to ensure safe use next time.