7.7 Weigh Scale verification
This operation is for checking the output of the weigh sensor . If it does not
work correctly, an additional calibration is needed.
You need something with a known weight .for example prepare a 200 g
standard sample weight.
1. Make sure the tank is fixed on the scale properly without any disturbance or
leaning against.
You can use
button to choose which wight sensor for checking ,for
example,we choose the sensor of the tank,press
button to confirm