Installation
Interfaces
to
connect
external
devices
The
following
devices
can
be
connected
to
the
ports
Com1
(
5
,
Fig.
6)
and
Com2
(
4
,
Fig.
6):
-
VISULENS
500
Digital
Lensmeter
from
software
version
v1.2.3
-
VISUREF
100
Autorefractometer/Keratometer
from
software
version
v2.1.7
Proceed
as
follows:
•
Please
turn
all
of
the
equipment
involved
off.
•
Connect
the
data
interfaces
for
the
devices
listed
above
with
the
connections
in
the
junction
box
to
be
used
for
this
purpose.
In
this
regard
please
use
the
interface
cable
supplied
and
select
the
following
allocation:
-
VISULENS
500
to
interface
Com1
-
VISUREF
100
to
interface
Com2
•
Attach
both
ends
of
the
interface
cable
to
the
respective
device
and
tighten
the
screws
on
the
cable
plugs
and
cable
sockets.
•
Please
ensure
that
the
interface
parameters
for
the
external
devices
have
been
reset
to
their
factory
settings.
Interface
to
VISUSCREEN
control
unit
The
interface
to
the
VISUSCREEN
control
unit
is
a
wireless
connection
using
Bluetooth
technology.
The
junction
box
broadcasts
and
receives
data
via
the
antenna
(
5
,
Fig.
18).
The
Bluetooth/USB
adapter
(
3
,
Fig.
18)
in
the
USB
interface
for
the
VISUSCREEN
control
unit
is
employed
as
the
remote
station.
Proceed
as
follows:
•
Loosen
the
knurled
screw
(
1
,
Fig.
18)
on
the
underside
of
the
housing.
•
Remove
the
cover
(
2
,
Fig.
18)
by
pulling
it
gently
forward,
and
then
upwards
out
of
the
retaining
tabs
of
the
housing.
•
Plug
the
Bluetooth/USB
adapter
into
a
free
USB
port
in
the
left
upper
corner
of
the
unit
(
3
,
Fig.
18).
31