Standard 25
ICS
B 40-029 e 01/98
3-7
Complete phase contrast is available only if the dark
phase ring in the objective covers the bright phase
stop in the condenser completely and precisely (also
see Fig. 3-7/
B
).
To achieve this, the diopter or the centering
telescope (3-6/
3
) must be inserted in tube (3-6/
4
)
instead of an eyepiece. Focus on the phase ring by
turning the eyelens (3-6/
1
) of the centering
telescope via the knurled ring (3-6/
2
).
Equipment B
phase contrast condenser II Z 0.9 Ph1, 2, 3
(3-8/
1
)
phase contrast objectives labelled "Ph 2"
(3-6/
11
), suitable also for transmitted-light
brightfield
Settings:
Insert phase contrast condenser II Z 0.9 Ph 1, 2,
3 (3-8/
1
) in the condenser carrier as described in
section 2.
Screw phase contrast objective (3-6/
11
) into the
nosepiece and swing it into the optical beam
path.
Set the object in brightfield first by having the
turret disk (3-8/
5
) click-stop in position "J".
Swing in the condenser front lens (3-8/
2
) via
lever (3-8/
4
).
Reduce luminous-field diaphragm (3-6/
5
), focus
it using the condenser, center it and open it
again (as described in section 3.2, page3-5).
Optimum image contrast in brightfield is
obtained by setting the aperture diaphragm via setting wheel (3-8/
3
). The aperture diaphragm can be
centered using knurled wheel (3-8/
6
) and fixed using lever (3-8/
7
) which can be clamped in position.
Swing phase stop Ph 1, 2 or 3 assigned to the phase contrast objective in the beam path via turret
disk (3-8/
5
).
Check centering of both phase stops using the centering telescope (3-6/
3
) in the tube, as described
above.
If the phase rings are out of center, as shown in Fig. (3-7/
A
), use knurled wheel (3-8/
6
) to make the
bright phase stops in the condenser exactly congruent with the dark phase ring in the objective, as
shown in Fig. (3-7/
B
).
Fig. 3-7
Centering of phase stops in phase
contrast
6
Fig. 3-8
Setting of phase contrast using Ph
condenser II Z 0.9
Содержание Standard 25 ICS
Страница 1: ...Standard 25 ICS Transmitted Light Microscope Operating Manual...
Страница 22: ...Standard 25 ICS 2 2 B 40 029 e 01 98...
Страница 30: ...Standard 25 ICS 3 2 B 40 029 e 01 98...
Страница 46: ...Standard 25 ICS 4 2 B 40 029 e 01 98...
Страница 56: ...Standard 25 ICS 4 12 B 40 029 e 01 98...
Страница 58: ...Standard 25 ICS A 2 B 40 029 e 01 98...
Страница 61: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...