Operation
000000-1363-329 SL 120 11.07.2006
24
Power supply
•
Actuate the power switch (
1
, Fig. 5) of the power supply to switch it
on.
•
Afterwards, turn on the brightness control (
2
instrument base. No slit light in the left final position, in the right
final position corresponds to the maximum illumination intensity.
Patienten positionieren
•
Adjust the patient's seat height as required.
•
If possible, adjust the table height in such way, that the patient is
slightly leaning forward. Thus, a stable rest of the patient's chin and
forehead is achieved. For the same reason, one may move the
patient's seat slightly away from the instrument or set it higher.
•
Adjust the chin rest (
6
, Fig. 23) vertically until the patient's eyes are
approximately level with the red ring marks (
7
•
Turn the joystick (
1
, Fig. 23) to set the slit lamp to mid-position of the
vertical adjustment range (red mark (
3
General notes on instrument operation
•
Choose a slit length that only illuminates that section of the eye
which is to be examined. Otherwise, sections of high reflectance (e.g.
sclera) may be illuminated, thus causing disturbing glare.
•
Turn the filter wheel (
4
, Fig. 23) to choose a filter as required.
•
For retroillumination, set the illumination angle as small as possible
and slightly decentre the illumination by means of control (
5
•
When using the tonometer, we recommend one looks with the right
eye through the left eyepiece while observing the patient with the
left eye.
•
The use of either convergent or parallel viewing path depends on
your choice.
Fig. 23
Adjust slit lamp
1
2 3
4
5
6
7