Care and Troubleshooting 

Primo Star 

 

 
 
26 

B 46-0071 v   03/06 

En
g

lish 

4.2 Troubleshooting 

Problem Cause 

Remedy 

Nosepiece with objective has not been 
switched into click-stop position. 

Switch nosepiece with objective into 
click-stop position. 

Condenser has not been set correctly.  Set condenser correctly. 

The aperture diaphragm has not been 
adjusted correctly. 

Adjust aperture diaphragm correctly. 

The luminous-field diaphragm has not 
been adjusted correctly. 

Adjust luminous-field diaphragm 
correctly. 

The field of view is not 
completely visible. 

The filter has not been inserted 
correctly in the filter mount. 

Insert filter correctly into filter mount. 

The aperture diaphragm has not been 
opened to the correct size. 

Set opening of aperture diaphragm to 
correct size. 

Condenser not focused correctly. 

Focus the condenser. 

Wrong cover slip thickness selected 
for use of transmitted-light objectives 
corrected for 0.17mm cover slips. 

Use standard 0.17 mm cover slips. 

Use of no or non-specified immersion 
oil with immersion objective 

Use the supplied immersion oil. 

Air bubbles in immersion oil 

Remove the bubbles by applying new oil 
or moving the objective to and fro. 

Immersion oil on the front lens of a 
dry objective. 

Clean the front lens of the dry objective. 

Low resolving power, poor 
image contrast 

Dirt or dust on the optical surfaces of 
objectives, eyepieces, condenser, or 
filters 

Clean the respective optical 
components. 

Greater focus differences after 
changing the objective 

The focusing eyepieces have not been 
set correctly. 

Set the focusing eyepieces to the 
corresponding defective vision. 

Power plug has not been plugged into 
power outlet. 

Connect power plug to power outlet. 

The 

6V 30W

 halogen lamp or 

the LED source does not light 
up although the microscope 
has been switched on. 

6V 30W 

halogen lamp or LED source is 

defective. 

Replace the defective 

6V 30W

 halogen 

lamp or the LED module 

End of average service life of 6

V 30W

 

halogen lamp has been reached. 

Replace the 6V 30W halogen lamp.  

Power cable Incorrectly installed or 
broken power cable 

Connect the power cable correctly or 
replace it. 

The pins of the 6V 30W halogen lamp 
have not been inserted correctly in the 
lamp holder. 

Insert the pins of the 6V 30W halogen 
lamp correctly in the lamp holder. 

The 

6V 30W 

halogen lamp is 

flickering, its light intensity 
unstable, the illumination 
inhomogeneous 

The pins of the 6V 30W halogen lamp 
have not been inserted symmetrically 
in the lamp holder.  

Insert the pins of the 6V 30W halogen 
lamp symmetrically in the lamp holder. 

Stage comes down by itself, 
image focus is unstable 

Adjusted torque of coarse focusing 
drive is too low. 

Increase the torque of the coarse 
focusing drive to make motion stiffer. 

 

Содержание Primo Star

Страница 1: ...Primo Star Operating manual Bedienungsanleitung Mode d emploi Instrucciones de manejo...

Страница 2: ...sent mode d emploi Nous vous prions par cons quent de lire attentivement les informations contenues dans ce document et de respecter notamment les consignes relatives la s curit d utilisation Le fabr...

Страница 3: ...ry distance and viewing height 16 3 3 2 Compensating for ametropia and inserting the eyepiece pointer 16 3 3 3 Adjusting transmitted light brightfield on the Full K hler microscope 17 3 3 4 Adjusting...

Страница 4: ...a possible hazard to the instrument or system CAUTION Disconnect the plug in power unit from line power before opening the microscope NOTE This symbol refers you to advice that you must observe under...

Страница 5: ...icroscopes should not be disposed of with household waste dispose of them in compliance with the provisions of the law 1 5 Use The microscope including its original accessories must not be used for mi...

Страница 6: ...by qualified specialists or service staff The operation of the instrument in explosion risk environments is not allowed Make sure to read the safety notes provided with Immersol 518 N immersion oil I...

Страница 7: ...ered Possible defects must be notified to us immediately and steps be taken to minimize damage If notified of such a defect the manufacturer is obligated to rectify it at his discretion either by repa...

Страница 8: ...Carl Zeiss Description Primo Star 8 B 46 0071 v 03 06 English 2 DESCRIPTION 2 1 System overview...

Страница 9: ...Primo Star Description Carl Zeiss B 46 0071 v 03 06 9 English...

Страница 10: ...Plug in power unit 6 Stand 7 ON OFF switch illumination intensity 8 Illumination module 9a Luminous field diaphragm adjustable 9b Luminous field diaphragm fixed 10a Condenser carrier Full K hler 10b...

Страница 11: ...ogen lamp LED illumination or illuminating mirror Integrated fixture for external power supply and cable incl cable with multi standard plug and country specific plug adapters Plastic coated carrying...

Страница 12: ...Phase contrast objective with green inscription and phase stop Ph 2 Objective magnification multiplied by eyepiece magnification results in overall visual magnification e g 10 x 10 100x Numerical aper...

Страница 13: ...specific adapters Fig 3 4 if necessary To this end pull off the attached adapter and plug on the desired adapter Connect the plug in power unit to a power outlet If the plug in power unit cannot be p...

Страница 14: ...Carl Zeiss Start Up and Operation Primo Star 14 B 46 0071 v 03 06 English 3 2 Controls Fig 4 Controls...

Страница 15: ...a Knurled ring for adjustment of luminous field diaphragm with Full K hler equipment only 10b Luminous field diaphragm fixed in Fixed K hler equipment 11a Condenser entering screws in Full K hler equi...

Страница 16: ...Fig 7 3 are equipped with fold over rubber eyecups Fig 7 1 pulled out Fig 7 2 folded over Both eyepieces are suitable for spectacle wearers Additionally they contain a focusing ring for the compensat...

Страница 17: ...the slide by means of the spring lever Fig 4 12 If the microscope stand is equipped with a phase or dark filed slider pull this slider out to the left as far as it will go Turn rotary knob Fig 8 1 on...

Страница 18: ...e Specimen field size and objective aperture change after every objective change Therefore repeat the adjustment of luminous field diaphragm and aperture diaphragm to obtain optimum results 3 3 4 Adju...

Страница 19: ...n Push the slider to the right as far as it will go to move the phase stop into the light path Move the control lever of the aperture diaphragm to fully open it on the Full K hler model or set it to t...

Страница 20: ...4 in the stand Then place the tube horizontally onto the stand In doing so the groove on the underside of the tube must be located above the third holding element Fig 11 3 of the stand Turn the tube...

Страница 21: ...ngers as this will reduce lamp life If the stand is equipped with an LED source replace the complete illumination module including the LED Fig 13 1 Push the illumination module Fig 13 1 or 2 back into...

Страница 22: ...sert Fig 15 1 or Fig 16 3 Push the condenser carrier backward against the spring and inclining it remove it upward from the condenser carrier Fig 15 4 Insert the mirror Fig 16 5 from top through the a...

Страница 23: ...rst install the photo tube Fig 17 7 if necessary refer to Section 3 4 1 Loosen clamp screw Fig 17 8 and remove the dust cap from the photo tube Installing a compact digital camera Sliding mount Fig 17...

Страница 24: ...Loosen grub screw Fig 17 5 Vary the camera lens lens mount distance gradually i e displace the sliding mount with camera on lens mount in defined steps Zoom the camera lens through from wide angle W...

Страница 25: ...owing on it and subsequent wiping it off with a cloth moistened with water to which you may also add a mild detergent Wipe off stubborn oily or fatty dirt immersion oil fingerprints with a cotton bud...

Страница 26: ...jective Low resolving power poor image contrast Dirt or dust on the optical surfaces of objectives eyepieces condenser or filters Clean the respective optical components Greater focus differences afte...

Страница 27: ...ure 5 to 40 C max 75 at 35 C 800 hPa to 1060 hPa Operating data Protection Class Protection Type Electrical safety Pollution degree Overvoltage category Radio interference suppression Line voltage Lin...

Страница 28: ...PL 10x 20 Br foc Specimen stage Dimensions width x depth Stage travel X x Y Coaxial drive Vernier scales Specimen holder Mechanical stage 75x30 right left 140 x 135 mm 75 x 30 mm Optionally right or l...

Отзывы: