Installation Instructions
Page
4
2. Selection of installation site and conditions
When selecting the installation site, make sure that the conditions specified below either already have been met or are
subsequently initiated by you.
The system includes the following components:
1. METROTOM
– Full protection enclosure
– Positioning unit
– Flat bed detector
– Microfocus X-ray tube
– Rotary table
2. Controller
3. Data station (peripheral devices)
2.1
Required room height
This dimension is the sum of the height of the METROTOM and the minimum clearance below the ceiling.
Example
:
2.2
Required floor space
A cut-out floor plan is attached to help you gain a quick overview of your floor space requirements. When drawing up
the overall plan, please make sure that your METROTOM is accessible from all sides.
Select the installation site so that the METROTOM is located at least 700 mm away from the wall.
2.3
Transport paths
Check the floor and ceiling loading capacity of all transport paths leading to the installation site. The doorways and trans-
port paths must be at least 10 cm wider and higher than the METROTOM assemblies to be transported, including the
transport vehicle. The dimensions and weights of the METROTOM assemblies are specified in this text.
2.4
Ambient conditions
MetroTomography
®
systems are high-tech products. Their specified precision and quality can be attained only if the fol-
lowing conditions are met.
Permissible storage temperature range
5°C to 40°C
Pay special attention to the storage temperature during the winter. Storage at temperatures below 5
o
C must be avoided.
Ambient temperatures required for operational readiness
Permissible ambient temperature: +15°C to +35°C
Permissible air humidity:
40% - 70% without condensation
Air humidity values outside of the permissible range require an additional air conditioning unit.
Height of METROTOM .................................................................................. 2150 mm
Minimum clearance below ceiling ................................................................... 800 mm
Required minimum room height above floor or foundation ................. 2950 mm