Automatic pinhole adjustment:
The automatic adjustment allows the pinholes to
be used with any combination of beam splitters
and more than one active channel.
Please do not make any program
manipulations while the automatic
pinhole adjustment is running (status
display is red - busy).
•
Click on the
Adjustment
Adjust
Automatically
button.
−
The
Requirements for Adjustment
window will then appear.
•
Meet the requirements listed in the
Requirements for Adjustment
window
(Fig. 559). Press
Cancel
to abort the automatic
pinhole adjustment. Press the
OK
button to
proceed.
−
Pinhole adjustment will then run
automatically and the Scan window will
appear (see Fig. 560). The adjusting
procedure takes approx. 3 min.
−
The determined data are stored
automatically and will be available for all
further examinations using the same
configuration.
A change of the pinhole diameter made
manually in the
Pinholes section
panel
will update the pinhole diameter in the
Channels
tool and vice versa.
Pinhole adjustment should be done on a
regular basis (monthly, depending on
system usage).
Fig. 559
Requirements for Adjustment
window
Fig. 560
Pinhole adjustment scan window
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......