Using  the  optimal  distance  between  slices  ensures  the  best  coverage  and  3D 

reconstructions,  while  avoiding  excessive  bleaching  and  photo-toxicity  on  your  sample  (see 

Nyquist  sampling).  Briefly,  if  you  use  a  step  size  smaller  than  what  the  Nyquist  sampling 

theorem  recommends  you  will  not  gain  any  more  information  and  you  will  likely  increase 

bleaching  and  photo-toxicity on your sample  by  imaging the  same “information” more  times 

than needed. You can choose the z-range to acquire by setting the first and last frames of the 

stack or set the current frame as the central frame of the pile and choosing how many slices 

you  want  to  acquire.  If  you  are  using  definite  focus  using  Center  might  be  the  best  option. 

Also, when using Center insert an odd number of slices so the current frame is acquired. 

  

Time Series 

This allows you to perform time-lapse experiments, working best with Definite Focus in 

longer  time-lapses  to  ensure  the  sample  stays  focused.  Number  of  cycles  is  the  number  of 

pictures  to  take,  if  zero  is  selected  the  software  will  take  pictures  until  manually  stopped. 

Interval is the time between each cycle, note that the interval timer resets when the picture 

begins being taken, if the interval is shorter than the acquisition time the next cycle begins as 

soon as the previous ends. 

 

Tile Scan 

Using Tile Scan you can have larger samples that do not fit in a single frame in the same 

image  with  high  magnification.  This  takes  multiple  images  that  overlap  each  other  (at  least 

10% but optimally 30%) and then they can be stitched together into a single image. There are 

three modes of tile scanning: 

-

 

Centered  Grid

  -  uses  the  current  frame  as  a  central  point  and  acquires  it  and  the 

area around it; 

-

 

Bounding Grid

 – you set the top left corner and bottom right corner of the area you 

want to acquire; 

-

 

Convex Hull 

– is used for irregular samples where a normal rectangular area would 

have  many  blank  frames  that  don’t  need  to  be  acquired,  therefore  reducing 
acquisition time.  

Содержание LSM 880

Страница 1: ...Zeiss LSM 880 with Airyscan Confocal Light Microscopy Basic User Guide...

Страница 2: ...tors 7 Transmitted Light Techniques 7 Incident Light 7 Features and Specifications 7 Objective 8 Filter Sets for Incident Light 9 Additional Equipment 10 Computer 10 Monitor 10 Potential Techniques 10...

Страница 3: ...3 Smart Setup 17 Lasers 18 Imaging Setup 18 Acquisition Mode 21 Channels 22 Definite Focus 2 23 Multidimensional Acquisition 23 Z Stack 23 Time Series 24 Tile Scan 24 Positions 25 Regions 25...

Страница 4: ...350 nm Z even in thick samples The introduction of a Fast module for the Airyscan offers a satisfactory compromise between perfect optical sections and live cell imaging experiments The Fast Airyscan...

Страница 5: ...enide phosphide GaAsP detector a 63x Plan Apochromat 1 4NA DIC oil immersion objective Zeiss and the 405 nm 488 nm and 561 nm laser lines The Zeiss Zen 2 3 black edition software was used to control t...

Страница 6: ...pment cannot be used without official training Before your first use contact Mariana Ferreira to schedule training Save all data into your own storage device as we routinely clean the computer s data...

Страница 7: ...ast DIC Phase Contrast Bright Field Incident Light Fluorescence HXP UV lamp Features and Specifications Axio Observer 7 SP Inverted Microscope Stand Motorized Stage 130 x 100 STEP Stage Controller XY...

Страница 8: ...nterference Contrast DIC II Differential Interference Contrast DIC III Lateral Resolution XY 525 nm m 2 0 7 0 2625 0 15 Nyquist Sampling XY 525 nm m 3 0 28 0 105 0 06 Zoom to Nyquist Sampling XY 525 n...

Страница 9: ...he Locate Tab Emission Filters for Airyscan and Fast Airyscan Position Wavelengths Emission Colours Optimal Fluorophores 2 BP 420 480 BP 495 550 Blue Green to Yellow DAPI GFP FITC YFP 3 BP 420 480 BP...

Страница 10: ...12 free slots 4x 2TB 4 TB RAID 10 NVIDIA Quadro K2200 4GB OS Windows 7 Ultimate Embedded x64 Available Software Zeiss Zen 2 3 Monitor Brand HP Model DreamColor Z32x Diagonal 32 Native Resolution 4K UH...

Страница 11: ...plore your fluorophores Learn their excitation and emission wavelengths and what are their optimal laser lines and filters This is necessary to use the confocal to its full potential Sample Preparatio...

Страница 12: ...see the Windows start up screen 5 Turn on the Components Switch 6 Wait for everything to finish initializing 7 Open Zen Black on the computer 8 Choose Start System and wait for the initialization to...

Страница 13: ...laser in Standby mode and turn all other lasers off b Otherwise turn all lasers Off 5 If turned Off it will take the Argon Laser at least 5 minutes to cool fans turn off 6 Shutdown the computer and th...

Страница 14: ...menu bar above and a system information bar below The Image and Document Manager Section shows all the currently open files and allows you to save and close the currently selected image The Image Disp...

Страница 15: ...transmitted light and off The transmitted light setup is used for every brightfield configuration with the type brightfield phase contrast or DIC being dependant on the condenser setting and prisms o...

Страница 16: ...live feed Snap Acquire current plane with your settings Additional available acquisition options check to show options tab To acquire images with these extra options you must use Start Experiment ins...

Страница 17: ...ee possible regular confocal configurations are generated with optimization for speed Fastest for Best Signal and a compromise between both Smartest Each configuration will display the speed percentag...

Страница 18: ...d The Argon laser takes a few minutes to warm up Imaging Setup At the top of this tab you have Tracks There can be up to four tracks simultaneously each with their own imaging mode and lightpath setup...

Страница 19: ...or strong fluorophores with Ch3 detector having an extra sensitivity to the red part of the spectrum 5 5 You can also identify the fluorophore from the dropdown list like in Smart Setup and assign its...

Страница 20: ...20 2 3 4 7 8 5 6 9 1...

Страница 21: ...Airyscan requires Optimal frame size Higher speeds give noisier images The recommended speed is 5 or 6 Check the Direction is bi directional as it is faster Use either 8 Bit or 16 Bit for Bit Depth p...

Страница 22: ...t to the histogram in the display area to see if you need to adjust the laser power If you are using the full dynamic range a lot of RED pixels means you are saturating the detector all max intensity...

Страница 23: ...g media such as Vectachield will be suboptimal unless diluted with an aqueous buffer To use this first select Definite Focus in the Focus Devices and Strategy tab then in the Definite Focus tab click...

Страница 24: ...te Focus in longer time lapses to ensure the sample stays focused Number of cycles is the number of pictures to take if zero is selected the software will take pictures until manually stopped Interval...

Страница 25: ...of the current frame will be saved Regions The pixel by pixel image acquisition of the confocal allows you to acquire only Regions Of interest ROIs instead of the whole frame You can make ROIs using...

Отзывы: