13
Shutdown
1.
Save your images in Users folder in the D drive;
2.
Close Zen Black;
3.
A window to save any unsaved images/profiles should appear if there are any,
DO NOT
overwrite other users’ experiment profiles!
4.
A Laser Off window appear:
a.
If someone is using the system in the next
4 hours
leave the Argon laser in
Standby
mode and turn all other lasers off;
b.
Otherwise turn all lasers
Off
.
5.
If turned
Off
it will take the
Argon Laser
at least
5 minutes
to cool (fans turn off);
6.
Shutdown the computer and then wait for the
Argon Laser
fans to turn off;
7.
AFTER
the Argon Laser cools down:
a.
Turn off the
Components Switch
;
b.
Turn off the
Systems/PC Switch
;
c.
Turn off the
Main Switch
;
8.
Put the dust cover back on.
ATTENTION
Turning the switches off before the Argon Laser module is done cooling can damage the
laser!