Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
Troubleshooting
M-3
Video Eye Monitor
Disappeared
• Press EYE button (
)
• Eye monitor is only present on some screens (Start of Test, Testing, etc.)
Too dark or too light
• Press “+” to lighten or “-” to darken (
Time and Date
Time display on screen is incorrect
• Correct by pressing DATE AND TIME on System Setup screen (
)
Date on screen is incorrect
• Change by pressing DATE AND TIME on System Setup screen (
)
Test date is incorrect on printout
• Change by pressing CHANGE PATIENT DATA on File Functions screen (
Printer
Does not print
• No power to printer (no lights visible on front panel) (
)
- Check to ensure power cord is attached and check power switch (
)
- Check power to power table (
• No paper in printer (
)
No signal from HFA II-
i
• Check printer cable connection (
• Check cable is correct and not damaged.
• Check to ensure correct printer is selected on System Setup menu (
• Printer buffer full - turn off printer and turn back on.
• No printout selected to print (
).
• Other printers: Consult specific printer manual.
Improper printout
• Random characters or multiple pages printed.
- Check that correct printer is selected on System Setup menu (
).
- Printer needs to be turned off to be reset.
• Abnormal value printed - reprint visual field.
(printer occasionally prints double characters).
• Printrex printout blank - paper may be upside-down.
Printouts fade over time
• Do not store printouts in plastic covers or use adhesive tape on thermal paper
(
).
Printout Problems
No Single Field Analysis printout
• Test pattern not eligible for STATPAC analysis (
).
• One or more test parameters not valid for STATPAC (
Quad totals not shown
• Available on Three-in-One printout only (
No SF or CPSD values on threshold printout
• SF and CPSD are not displayed when SITA test strategy is used (
• Fluctuation was turned off for the test (CPSD is not calculated when fluctuation is
off) (
Glaucoma Hemifield Test (GHT) results not
displayed
• FastPac test strategy used [GHT not available with FastPac] (
)
• Test not eligible for STATPAC analysis (
).
Table M.1 Troubleshooting Table
Problem
Possible Cause or Solution
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......