Microscope Illuminating System
Start-Up
HBO 100
Inserting the Collector Attachment
Carl Zeiss
B 46-0063 e 03/06
17
3.6 Inserting the Collector
Attachment
1 Collector attachment
2 Spring ring
3 Heat protection filter
4 Clamping ring
Fig. 11 Inserting the collector attachment
•
Pull out clamping ring (Fig. 11/
4
).
•
Remove heat protection filter (Fig. 11/
3
).
•
Screw the collector attachment into the
lamp as far as it will go (Fig. 11/
1
). The
retaining ring on the collector attachment
must point to the outside.
•
Clamp the collector attachment by inserting
the spring ring (Fig. 11/
2
). The spring ring
engages in the groove of the lamp.
•
Insert heat protection filter and lock it by
means of the clamping ring.