Start-Up
Microscope Illuminating System
Carl Zeiss
Lamp Insertion and Change
HBO 100
12
B 46-0063 e 03/06
Procedure:
•
Switch off the transformer.
•
Disconnect the microscope illuminator cable
from the transformer.
•
Allow the microscope illuminator to cool
down for 15 minutes.
•
Wear protection mask and safety gloves.
•
Remove microscope illuminator with cooled-
down lamp from the microscope.
•
Turn control knob (Fig. 4/
5
) to move
collector to mid position.
•
Turn both fixation screws (Fig. 4/
2
) in the
direction indicated by the arrow as far as it
will go; remove top part of lamp upward.
•
Operate clamping lever (Fig. 4/
7
) at heat sink
(Fig. 4/
1
) and withdraw heat sink upward
from the safety holder (in commissioning) or
from the lamp (Fig. 4/
6
).
•
Put heat sink down beside the illuminator.
For this, you need not disconnect the electric
connection to the heat sink.
•
Push clamping lever at lamp base (Fig. 4/
4
)
down, to release the lamp lock;
remove the defective lamp or the safety
holder (Fig. 4/
6
) from the lamp mount
(Fig. 4/
3
) in the bottom part of the
illuminator. Release the lever.
•
Dispose of the defective lamp in accordance
with the relevant regulations
(see section
1.7 – Hints on Disposal
).
•
Push clamping lever at lamp base (Fig. 4/
4
)
down and insert the new lamp into the
mount in the bottom part of the illuminator
(as far as it will go); release the lever to
clamp and center the lamp.
Do not exert any force when inserting the
lamp.
•
Take the heat sink (Fig. 4/
1
), push down its
clamping lever (Fig. 4/
7
) and put the heat
sink onto the lamp (Fig. 4/
6
). Release the
clamping lever. By doing so, the electric
connection is established and the heat sink
clamped.
Do not exert any force to the lamp, when
attaching the heat sink.
•
Put the top part of the illuminator onto the
bottom part.
•
Turn both fastening screws (Fig. 4/
2
) on the
bottom part of the illuminator oppositely to
the direction indicated by the arrow as far as
it will go.
•
Turn the control knob of the collector back
to mid position, if necessary.