Microscope Illuminating System
Start-Up
HBO 100
Focusing and Centering
Carl Zeiss
B 46-0063 e 03/06
15
3.4 Focusing and Centering the
HBO 100 Microscope
Illuminator, self-adjusting
(423011-9901)
1 LED key
2 Collector control knob
Fig. 9
Adjustment of HBO 100, self-adjusting
Automatic adjustment of the HBO 100 requires
the collector to be set to mid position.
•
Turn the collector control knob (Fig. 9/
2
)
into mid position.
After the transformer of the connected
HBO 100 microscope illuminator was switched
on, the five minutes warm-up phase begins.
The LED key (Fig. 9/
1
) lights orange during
warm-up.
After the five minutes are over, automatic
adjustment is started.
The LED key flashes green during the
adjustment.
The adjustment is finished after about 40
seconds. The LED key lights permanently green
then.
If the adjustment failed, the LED key lights red.
•
In this case, you can start automatic
adjustment manually by a short pressure
(approx. one second) of the LED key
(Fig. 9/
1
).
If the adjustment fails again, check whether
the lamp has been inserted correctly in the
base of the HBO 100 (pushed in as far as it will
go). For this, refer to
Section 3.1
.