DuraMax Installation site requirements
18
Securing the shopfloor base
To ensure the stability of the CMM, the shopfloor base must be anchored to the floor.
The fastening material for the base is included with delivery.
Note:
The boreholes to anchor the base must be drilled by the customer for warranty reasons.
330
660
920
732
1
965
Position of the boreholes for the shopfloor base (top view), M 1:10, dimensions in millimeters
Three fastening brackets (1) and three plug screws are included with delivery for fastening the shopfloor base to the
floor (foundation). These plug screws must be securely screwed into the foundation.
Risk of tipping if not properly anchored
If the floor covering is too thick and the plug screws cannot be anchored sufficiently as a result,
there is a risk of tipping.
In this case, customers must provide sufficiently long plug screws.