Instructions for Use
4 Daily Startup
CLARUS 500
4.1 Turn on the Instrument
2660021165005 Rev. A 2017-08
39 / 144
4 Daily Startup
4.1 Turn on the Instrument
CAUTION!
Do not unplug the instrument during startup
or instrument damage may occur.
u
Make sure the instrument has completed its system checks be-
fore you unplug or remove power.
Each time you turn on the CLARUS 500, system diagnostics checks
that the system initialzed correctly and is running properly. If the di-
agnosic check detects any problems, an error or alert message dis-
plays. For more information about startup errors and alerts, refer
to: Troubleshooting [
NOTE
If a critical system test fails
take the device out of service.
u
Contact your ZEISS representative.
u
If a non-critical test fails or a warning appears, the device can
still be used. Contact your ZEISS representative. (See Mainte-
nance [
NOTE! A System Administrator must assign an account for
each system operator.
1
2
Action
1. Turn on the instrument (green switch).
2. When the system test sequence completes, click
Continue
.
3. Log in.
ð
The
Patient
window displays.
Содержание CLARUS 500
Страница 1: ...CLARUS 500 Instructions for Use ...
Страница 98: ...Empty page for your notes ...
Страница 122: ...Empty page for your notes ...
Страница 124: ...Empty page for your notes ...
Страница 130: ...Empty page for your notes ...
Страница 132: ...Empty page for your notes ...
Страница 134: ...Empty page for your notes ...
Страница 136: ...Empty page for your notes ...
Страница 143: ...Instructions for Use Index CLARUS 500 2660021165005 Rev A 2017 08 143 144 Z Zoom 31 Zoom In Out Annotations 83 ...